STANDARD LIST OF EQUIPMENT FOR
ESTABLISHING/UPGRADING OF FORENSIC
SCIENCE LABORATORIES
Copyright ©DFSS, New Delhi, 2020
All rights reserved. No part of this publication may be reproduced, distributed or transmitted in any form
or by any means, including photocopying, recording or other electronic or mechanical methods, without
the prior written permission of copy right owner, expect in the case of brief quotations embodied in
critical reviews and certain other non-commercial uses permitted by copyright law.
First Edition: July 2020
Page 5 of 95
Page 6 of 95
Page 7 of 95
INDEX
S. No
Chapter/Content
Page Number
1.
Introduction
9
2.
Forensic Chemical Sciences
10-11
3.
Forensic Biological Sciences
12-13
4.
Forensic Physical Sciences
14-15
5.
Forensic Electronics (Cyber Forensics)
16-18
6.
Forensic Questioned Document Examination
19
7.
Forensic Investigation
20
8.
Specifications (Annexure 1-34)
22-80
9.
General Laboratory Consumables and Infrastructure
requirements (Appendix A-D)
82-91
Page 8 of 95
Page 9 of 95
INTRODUCTION
Forensic Science is the application of science which with its capabilities and facilities provides accurate
and timely results for the dissemination of justice speedily. The main aim of Forensic setup is to link the
crime with criminal on scientific grounds and thereby fair and transparent investigations and proceedings
of the case.
The constantly changing pattern in the execution of crimes has also posed a challenge to Forensic Science
Laboratories and necessitates the need to strengthen the laboratory examination practices. Directorate of
Forensic Science Services (DFSS) under Ministry of Home Affairs is playing a vital role to strengthen the
Forensic services in India by providing support to establish Hi-tech Forensic Science Laboratories and to
upgrade their human as well as technical resources.
However, DFSS often receives requests from various state Forensic Science Laboratories and other
stakeholders regarding basic requirements and infrastructure/equipment required to kick start the
laboratory.
Keeping in view the requirement of all the stakeholders, Directorate of Forensic Science Services, MHA,
Govt of India constituted a Committee by involving experts of Central and State FSLs vide letter No.
37(19)/2020-DFSS/6890-91 Dated June 5, 2020 (copy enclosed) to develop this manual with the sole aim
to provide assistance for the establishment /up-gradation of Forensic Science Laboratories in the country.
The forensic areas/fields addressed in this manual are broadly classified into six major groups which in
turn have subgroups of specialized branches.
1. Forensic Chemical sciences:
Forensic Chemistry
Narcotics
Toxicology
Explosives
2. Forensic Biological sciences:
Forensic Biology and Serology
Forensic DNA typing
3. Forensic Physical sciences:
Forensic Physics
Forensic Ballistics
4. Forensic Electronics:
Digital/Cyber forensics/computer forensics.
Audio and Video forensics
5. Forensic Questioned Document examination:
6. Forensic investigation:
Forensic Psychology examination
Crime scene investigation
This manual provides information regarding establishment/upgradation of cited fields in terms of major
and minor equipment. The specification of major equipment and checklist of general laboratory
infrastructure is annexed in the manual.
Page 10 of 95
1. FORENSIC CHEMICAL SCIENCES
1.1. Forensic Chemistry
Forensic chemistry deals with the chemical examination of liquors, identification of traces of fire
accelerants from burnt materials in bide burning and arson cases, detection of adulterants in petroleum
products, identification of corrosive chemicals in acid throwing incidents and identification of dyes in
bribe trap cases.
Types of Examination/Tests:-
Extraction of fire accelerant from burnt material/debris/partially burnt clothes/hairs etc.
Identification of residue of fire accelerant in extracted material.
Identification of Petroleum products by their hydrocarbon numbers.
Identification of dyes from hand wash/pocket wash in Bribe trap cases.
Determination of proof percentage and alcohol percentage in liquor cases.
Determination of lethal alcohols/ methanol in hooch tragedies.
Identification of gold in theft/burglary/fake ornament cases.
Identification of corrosive chemicals in acid throwing incidents.
1.2. Forensic Narcotics
The cases received in this branch are the street drugs seized by investigating officers. Majority of the
drugs are under NDPS Act 1985. The examination involves the identification and profiling of contraband
substances.
Types of Examination/Tests:-
Preliminary examination
Absolute /Fingerprint identification of drugs of abuse and their profiling.
Quantitative estimation of drugs of abuse.
1.3. Forensic Explosives
Explosives branch carries out the analysis of seized low and High explosive materials, analysis of post-
blast residues and pyrotechnics.
Types of Examination/Tests:-
Extraction of Post blast Explosive residue from Physical samples/debris.
Preliminary examination of Explosive substances and post-blast explosive residue.
Identification of Low explosives and their post-blast residue.
Identification of High explosives and their post-blast residue.
Determination of the composition of Pyrotechnics.
Reconstruction of Improvised Explosive Devices.
1.4. Forensic Toxicology
Toxicology branch undertakes the examination of biological/visceral samples for identification of poisons
in homicide, suicide and other unnatural death cases. The various types of samples received are stomach
with its contents, small intestine, a portion of the kidney, spleen lungs liver, brain blood, gastric lavage,
hairs, nails etc.
Types of Examination/Tests:-
Extraction of poisons from body fluids, viscera etc.
Preliminary examination
Qualitative and Quantitative estimation of Alcohol in blood samples
Determination of poisons and their metabolites in body fluids/gastric lavage /visceral
samples/liquid samples.
Page 11 of 95
The Major/ Minor equipment facilities required for such types of examination are:
S No
Major equipment
Proposed
Specifications
1.
Gas Chromatograph with an
Auto sampler
Annexure-1
2
UV-Visible
Spectrophotometer
Annexure-2
3
Gas Chromatography-Mass
Spectrometer with Auto
sampler
Annexure-3
4
Fourier Transform Infrared
Spectrometer ( FT-IR)
Annexure-4
5
High- Performance Liquid
Chromatography /Ultra-
high-Performance Liquid
chromatograph with auto
sampler
Annexure-5 & 6
6.
Ion Chromatograph
Annexure-7
7
Gas Chromatography
Headspace System (GC-HS)
Annexure-8
8
Automated Accelerated
Solvent Extraction System
Annexure-9
Minor Equipment
TLC Apparatus
Muffle furnace
FiltrationApparatus
Digital hot plate with the stirrer
Analytical balance
Tissue homogenizer
Digital Balance
Microscope
Microwave
Ultra sonicator
pH meter
Water bath
Thermometer
Fractional distillation apparatus
Hot air oven
Thin Layer Chromatographic Apparatus
Rotary Shaker
Micropipettes
Vortex mixer
Kozelka Hine Apparatus
Note: The materials and accessories required for spot tests/chemical tests/Density measurement etc. are
given in General Laboratory requirements Appendix
Page 12 of 95
2. FORENSIC BIOLOGICAL SCIENCES
2.1 Forensic Biology and serology
Biological and serological examinations are carried out to identify and characterize stains, hair, body
tissues, and microorganisms for species typing, grouping for individualisation and examination of body
fluids (semen, saliva, vaginal secretions etc.) for linkage purposes. Examination of insects in putrefied
bodies, diatoms and plant materials in drowning cases is also carried out in Biology section.
Types of examination/tests:
Location of stains and other microorganisms
Identification of characteristic type of stains
Identification of the origin of stains/ body fluids
Blood grouping
Identification of insects, Diatoms and plant materials in Drowning cases
Blood Pattern Analysis in Homicide and Murder cases
2.2 Forensic DNA Profiling
DNA analysis, also known as DNA profiling, DNA testing, DNA typing, or genetic fingerprinting,
involve the extraction, purification, amplification, sequencing and detection of unknown or suspected
DNA traces in a laboratory and their comparison with known/unknown DNA profiles to determine their
source.
Types of Examination/Tests:-
Extraction of DNA molecule from cellular material /other materials/Bacteria etc.
The quantisation of the DNA Molecule.
Amplification of extracted DNA molecule
Determination of DNA Sequence/profile
Comparison of questioned DNA Sequence with authenticated profile
Examination of paternity disputes
Examination of unidentified bodies by DNA typing
The Major/ Minor equipment facilities required for such types of examination are:
S No
Major equipment
Proposed
Specifications
1
Auto DNA Extraction
System
Annexure-10
2
PCR Thermal Cycler
Annexure-11
3
Real-Time PCR (RTPCR)
Annexure-12
4
DNA Sequencer
Annexure-13
Page 13 of 95
Minor Equipment
PCR Kits
Micro centrifuge
DNA Sequencing Kits
Vortex mixer
DNA Extracting Kits
Magnetic stirrer and stirring bars
Dry bath
Biological hoods and chambers
Temperature controlled Water bath
Digital Microscope
Analytical Balance
Refrigerators
Mechanical shaker
Anthropological Measurement tools and kits
Handheld Forensic Light Source with UV and
IR capability along with photographic filters
Stereomicroscope
UV sterilizer chamber
Gel Documentation System
Bio-safety Cabinet
pH Meter
Fume hood
Water Purification System
-80°C Deep Freezer
20°C Deep Freezer
Drying cabinet
Digital Autoclave
Tissue Lyser
Liquid handling System
Mini spin
Refrigerated centrifuge
Hot air oven
Vacuum pump
Mix mate
Thermo mixer
Gel electrophoresis system
Bone pulveriser
Microtome
Incubator
Note: The materials and accessories required for spot tests/chemical tests are given in General
Laboratory requirements Appendix A
Page 14 of 95
3. FORENSIC PHYSICAL SCIENCES
3.1. Forensic Physics
This branch conducts the examination of physical materials in cases of hit and run, burglary and
homicides etc. The samples involve are glass, soil, paints, Fibres, Electric failures, Cement and mortar,
Tool marks, tyre marks, Shoe Print analysis etc. The infrastructure required for the examination is:-
Types of Examination/Tests:-
Examination of imprints (tool marks/tyre marks/shoe marks/bite marks/cheiloscopy etc.)
Examination of soil samples.
Examination of ligature material such as wires, ropes, cords, cloth etc.
Examination of paint chips.
Examination of glass samples.
Analysis of short circuits/electric failures/component examination (circuitry concept, electronic
timers, electronic relay, solid-state device components) etc.
Analysis of cement and strength testing of building.
Examination of Counterfeit currency/security documents.
Examination of infringed products such as hologram stickers, lottery tickets, nano-bio products.
Restoration of erased/tempered serial numbers of vehicle/other items.
3.2. Forensic Ballistics
Ballistics examination involves the examination of firearms, ammunition and GSR particles. It includes
the determination of type, makes and calibre of firearm and ammunition and also to ascertain whether a
bullet or cartridge recovered has been fired from a suspected firearm. The examination also includes
identification of shooter based on GSR (Gunshot residue) particles recovered from target or suspect.
Types of Examination/Tests:-
Test firing for comparison
Comparison of questioned fired cartridge with a test cartridge
Comparison of questioned fired bullet with a test bullet
Gun Shot Residue ( GSR) Analysis
Trajectory Measurements
Reconstruction of a firing incident in Homicide/Suicide cases
Identification of Firearm (Factory made/country-made weapons) and parts thereof.
Serviceability/working condition of firearms/ammunition.
Identification of ammunition.
The Major/ Minor equipment facilities required for such types of examination are:
S No
Major equipment
Proposed
Specifications
1
Comparison of Microscope
Annexure-14
2
SEM-EDXA
Annexure-15
3
Ballistics Data Measurement
System
Annexure-16
4
X-ray Fluorescence
Spectrophotometer
Annexure-17
5
X-ray Diffractometer
Annexure-18
Page 15 of 95
6
Atomic Absorption
Spectrometer
Annexure-19
7
Inductively coupled Plasma-
Atomic Emission
Spectrometer (ICP-AES)
Annexure-20
Minor equipment
Bullet recovery Box
Hot Oven
Self-Protection firing system.
Furnace for 2000C
Wet lab for a chemical test.
Density Gradient Tubes of 6mm dia
Stereo Microscope.
Lighting board as a background for observation of
gradient tubes
Analytical Balance
Glassware in various sizes
Fire-arm mounting device.
Hot plates
Tool Box/Workshop Tools
Multi-meter
Chemical/Reagents/Glassware
Sieving machine
Viscometer
Sieves in different numbers
Sonometer
Burettes 25 and 50 ml
Polarising Microscope
Pipettes
Binocular Microscope
pH meter
Balance for weighing up to 100Kg
Tensile Testing Machine
Refractive Meter more than 1.7 RI
Mohes hardness digital testing tool
Weighing Balance
Tensile strength measuring equipment
Automated Ra-Top sieve shaker
High-end work station
Densitometer
Digital micrometre
Digital Vernier callipers
Digital dial callipers
Digital Tong tester
Digital tape
Digital SLR camera
Laser-based measuring tool (for measurement of
area/volume/distance)
Boroscope
Ultrasonic cleaner
Magneto-Optical device for the restoration
of erased numbers
Electrostatic Dust Print Lifter
Note: The materials and accessories required for spot tests/chemical tests are given in General
Laboratory requirements Appendix B
Page 16 of 95
4. FORENSIC ELECTRONICS
4.1. Digital/Cyber Forensics/Computer Forensics
This branch deals with the examination of digital material such as computers, Hard Discs, CDs, pen
drives, DVDs, mobile phone etc. confiscated by investigating agencies during the investigation. As
electronic media is fast changing, it is very much essential for any forensic laboratory to be equipped with
the latest art of Hardware and software required for seized samples.
Types of Examination/Tests: -
To recover the original data from the confiscated device for further analysis.
Write protection of data.
Imaging of digital evidence
Hash value verification
The Major/Minor H/w & S/w required for analysis of digital evidence is:-
S No
Major equipment
Proposed
Specifications
1.
Hardware imaging Device
Write Blockers Kit
Tableau duplicator
Falcon
Annexure-21
2
Software for Forensic
previewing, Imaging and
Analysis of Digital Media
(Encase software)
Annexure-22
3
Software for Forensic
Previewing, Image mounting
password cracking and
Forensic analysis of digital
media (Forensic Tool Kit6)
4
Forensic Tool Kit for
Incident Response of Digital
Crimes (VPER Kit)
5
Portable Image Device for
Incident Response
6
Forensic Triage Tool kit
7
Digital Forensic tool Kit for
the analysis of various
internet artifacts (Magnet
IEF)
8
Forensic software for the
analysis of MAC System
10
Software based tool for the
logical level analysis of
GSm/CDMA Mobile
phones
(UFED Link Analysis
Annexure-23
11
Hardware tool kit for
physical/logical level analysis
of GSM/CDMA Mobile
phones/SIM Cards
(UFED Touch-2)
Page 17 of 95
12
Mobile Phone Forensic tool
kit for logical analysis of
Chinese Mobile
Phones(*UFED Chinex)
13
Digital Forensic Tool Kit for
the physical imaging and
analysis of SIM Card
Memory
(Paraben)
14
Password Cracking tool for
Decrypting the passwords
protected files
Annexure-24
15
High configuration Server
and Workstations
Annexure-25
Minor equipment
Licensed Microsoft office and window
software
Computer systems
Safe Bag
UPS
Cyber Check Suite
Mac OS examiner
Autopsy
DVR Examiner
Other software for utilisation of acquisition of
data
CCTV Footage Acquisition Analysis(AMPED Five)
4.2 Audio and video analysis
This branch deals with the examination of Audio and video clips generated by using various media and
required for investigation purposes. It involves Data recovery, write protection, imaging and hash
verification for which high-end hardware and software is required.
Types of Examination/Tests: -
To recover the original data from the confiscated device for further analysis.
To compare the questioned Audio clip with Authenticated Audio clip
Voice Analysis / speaker identification & authentication
Analysis of Video clips/ CCTV clips and identify morphing
Authentication of audio & video clips
Identification of Source camera Comparisons of a questioned digital photograph with digital
photographs obtained from the questioned camera
S No
Major equipment
Application /Uses
Proposed
Specifications
1.
Software
1.Gold Wave
2.Multi speech
3.Audio Authentication and
Enhancement
Hardware
1. Analog to Digital converter
Hardware and software for
Audio Analysis
Annexure-26
2
Software
1. Video Authentication and
Enhancement
2.Video Editing
Hardware
Hardware and software for
video Analysis
Annexure-27
Page 18 of 95
1. High configuration Server
and Workstation.
3.
Facial Recognition Software
(3D)
For facial recognition/CCTV
footage analysis
Annexure-28
4
Image Analyser Software
Identification of Source camera
by comparison and examination
of sensor noise pattern in the
questioned digital image and test
shots
Annexure-29
Minor equipment
Licensed Microsoft office and window soft
wares
Computer systems
Microphones
UPS
Headphones
cables
Pinnacle software
DVR Examiner
Note: The materials and accessories required for setting up of cyber laboratory are given in General
Laboratory requirements Appendix C
Page 19 of 95
5. FORENSIC QUESTIONED DOCUMENT
EXAMINATION
The division carries out the analysis of questioned documents which involves cases of handwriting
comparison, Signature comparison, forgery of documents, ink analysis etc.
Types of Examination/Tests: -
Comparison of questioned Handwriting with Authenticated Handwriting
Comparison of questioned Signatures with Authenticated Signatures
Determination of addition/alteration/forgery.
Ink Analysis
Paper analysis
Determination of sequence of stokes
Deciphering of obliterated writing
Deciphering of indented writings.
The Major/ Minor equipment facilities required for such types of examination are:
S No
Major equipment
Proposed
Specifications
1
Multispectral Imaging system
( (VSC)
Annexure-30
2
High-resolution spectral
comparator (HRSC)
Annexure-31
3
High-End Computers with
server
Annexure-32
Minor equipment
High-end camera for photography of Documents
Stereo Microscope
ESDA For deciphering indented writings
Reprovit Unit for photography with SLR camera
and storage facility
Hand Magnifiers with/without a light source of
varying capacity: x3, x6, x8, x10
Examination Table with Transmitted Light box
U V LIGHT Source (wavelength-256nm,265nm,
313nm).
Anglepoise lamp
Page 20 of 95
6. FORENSIC INVESTIGATION
6.1 Forensic Psychology Division
This is related to the investigation and involves the extraction of truth from accused/ suspect by use of
scientific tools and methods.
Types of Examination/Tests: -
To extract the truth from the suspect/accused by the scientific manner
To determine the changes in voice when suspect/accused is lying and concealing information.
The equipment required to kick start the Psychology Division are;
S No
Major equipment
Proposed
Specifications
1
Polygraph
Annexure-33
2
Layered voice analyser
Annexure-34
3
High-End Computer with
server
Annexure-32
6.2 Crime Scene investigation
Majority of the times Forensic Scientists have to visit Scene of occurrences particularly in Explosion
cases, Cases of Arson and Burglary and accidents for collection of relevant physical evidence. The List of
minor equipment required to search the scene of occurrence is given in Appendix-D.
Page 21 of 95
Page 22 of 95
SPECIFICATIONS
Page 23 of 95
Page 24 of 95
Annexure 1
Gas Chromatograph (GC)
General Specifications: -
S No.
Specifications
1.
GC
Automatic microprocessor controlled with self- diagnostic
facilities
2.
OVEN
TEMPERATURE
Ambient to 450 O C or higher with an incremental rate of 50 O
C/minute or more with at least 6 or more ramp and fast cool
down of less than 10 minutes from 450 O C to 50 O C.
3.
Injector temperature
400O C or more
4.
Injector
Two injection ports split/ split less (one for capillary and one for
packed column-mounted simultaneously
5.
Flow controller for carrier
and carrier and detector
Automatic digital flow and a pressure controller/EPC/EFC/IEC
must be available
6.
Carrier gas
The system should have the facility to use helium and hydrogen as
the carrier gas in the GC system.
7.
Safety features
Auto leak detection of carrier gas and automatic shutdown of GC
in case of the leak
8.
GC system should have a gas saver mode
9.
Detector
Flame Ionization Detector-02 Nos The system should have two
nos of detectors (FID) mounted simultaneously.
FID-Temp Max-4000 C or more
Dynamic range 107
Minimum detection limit:2pgC/s
10.
Validation system
Total validation should be quoted for system and software IQ,
OQ or PQ
11.
Column
DB-1 MS or its equivalent 30 mX 0.32 mm(I.D) x0.25
micrometer film thickness- one number DB-5 MS or equivalent
30 m X0.25 mm(1.D)x0.25 micrometer film thickness-one
number.
3o/o SE -30 , OV-17 , Porapak g (1 each)
12.
Software
(with License)
Should be windows based operating system with validation
facility. The offered software should be suitable for the quoted
model GC along with performance validation.
The facility, multi component analysis, multilevel calibration,
baseline correction, methods of analysis, time programming,
plotting, graphical representation of the chromatograph and
storing of data
13.
Data station
The principal should supply the branded (DELL/HP/IBM etc)
computer system along with the licence of an operating system
with 21 branded LED monitor along with laserjet printer.
14.
All spare parts and accessories for the installation of the full system should be provided along
with one set of a tool kit.
15.
Spares and consumables-supplier/manufacturer should quote and supply essential spares and
consumables including septa (mention item wise) for two years month operation of the quoted
system.
16.
General conditions
I. The manufacturer and or supplier will provide a 2-year warranty from the date of
installation of the full GC system along with the supplied computer, printer and other
accessories for self-sustaining of the entire system.
II. Comprehensive and non-comprehensive AMC rate must be quoted for 5 years from
the post-warranty date. AMC includes Preventive maintenance- 4 visits per year every
quarter and breakdown visit as & when required.
Page 25 of 95
III. The manufacturer/supplier should certify that they will provide all the spares and
consumables at least seven years from the date of installation of the system.
IV. Training- The supplier should provide on-site operational/ application training for
laboratory scientists for at least 5 working days.
Page 26 of 95
Annexure 2
UV-Visible Spectrophotometer
General Specifications:-
1.
Optical System
Double beam
2.
Wave-length range
190 nm to 1100 nm
3.
Spectral bandwidth
0.2-3.0 nm or better
4.
Wavelength accuracy
± 0.1 nm in u.v. region or better.
± 0.3 nm in visible region or better.
5.
Measuring modes
Absorbance mode, % Transmission mode
6.
Light source
Pre-aligned Tungsten & Deuterium lamps
7.
Detector
High ending Photodiodes
8.
Operation mode
Stand-alone and PC control
9.
User-friendly UV-Vis
Software (with license)
The latest version of validated Windows-based UV-Vis software
with all licensed software to control and operate all functions of the
offered system with data acquisition facility along with performance
validation of the system.
10.
Data Station
The latest version of branded PC like DELL/HP/COMPAQ/
LENOVO/IBM makes along with the licence of the original system
with 21 LED monitor and laserjet printer.
11.
UPS
2 KVA or more with one hour back up with a full load of reputed
brand Such as Numeric/APC/Tata Libert
12.
Spares & consumables
Supplier/manufacturer should quote and supply essential spares and
consumables (mention item wise) for two years smooth operation of
the quoted system.
13.
General condition/Optional
a) The manufacturer and or supplier should provide a 2-year warranty from the date of
installation of the full system along with the supplier computer, printer, UPS and other
accessories for self-sustaining of the entire system.
b) Comprehensive and non-comprehensive AMC rate must be quoted for 7 years from the
post-warranty date (Non-comprehensive for first 2 years and comprehensive for next 5 yrs)
payable on year to year basis, Preventive maintenance-4 visit as and when required.
c) The manufacturer/supplier should certify that they will provide all the spares and
consumables at least for 10 years from the date of installation of the system. The supplier
should also certify that they will upgrade the software free of cost whenever such an
upgrade occurs.
d) The firm should supply tool kit, hard and soft copies of operation manual, and other
relevant literature along with a certificate of the license of the library, software and other
items wherever applicable.
Page 27 of 95
Annexure- 3
Gas Chromatography-Mass Spectrometer
General Specifications: -
S. No
For Gas Chromatography
1
GC
fully automated microprocessor controlled with extensive self-diagnostic
facilities and with Auto sampler
2
Auto sampler
T
he system should have auto sampler to adjust at least 100 vials
Injection cycle time should be less than 8 sec in specified conditions.
The Auto sampler has to support a wide range of injection volumes.
Auto sampler should be controlled by system software.
3
Oven
Temperature
450
0
C with 6 or more ramps facility settable from 0.1
0
C to 50
0
C/min or
more
4
Injector-
temperature
Up to 400
0
C in 1
0
C increment or more
5
Injector
spilt and spilt less injector for capillary column
6
Flow controller
for carrier
automatic digital flow and pressure controller /EPC
7
Safety features
auto leak detection of carrier gas & automatic shut down of GC in case of a
leak
8
GC system should have a gas saver mode
9
Software
Windows-based
10
Validation
GC and software should have a validation facility
11
Column
DB-5 MS/ HP 5 MS, 15 m X 0.25 mm x 0.25 micrometre one no.
12
ISO 9001 certification and GLP compliance
Mass Spectrometer
1
Mass
27pprox.27-
quadruple with pre-filter
2
Ionization
source-
electron ionization ( EI) and chemical ionization ( CI) ( Positive and
negative) equipped with advance non-vacuum breaking cleaning facility
3
Mass range-
10-1000 amu or more
4
Electron
energy/
ionization
voltage-
adjustable electron energy up to 150 eV or more
5
Tune facility -
should have auto & manual tune facility
6
Scan rate
variable up to 5600 amu/ sec or more
7
Vacuum pump
capacity
200 lit/ sec or more air-cooled turbo molecular pump
Page 28 of 95
8
Sensitivity/
detection limit
1 pgstd OFN scanning from 50- 300 u at m/z 272 with Helium as carrier
gas should produce S/N 400: 1 or more in EI mode and equivalent with
any other compound
10 pgstd BZP should produce S/N 10:1 or 100 pgstd BZP scanning from
80 to 230 amu in m/z 183 in methane gas should produce S/N 125:1 in
PCI mode or better and equivalent with any other compound
1 pgstd OFN should produce S/N 1000:1 or 2 ul injection of 100 fgstd
OFN should produce S/N 300:1 in NCI mode or better and equivalent
with any other Compound at m/z at 272 amu
9
Calibration
facility
should have built calibration facility with internal calibrating material with
NIST/ ASTM or equivalent traceability
10
scanning facility
simultaneous scanning in SIM and full scan (TIC) mode
11
Detector
Dynode/photomultiplier
12
Interface
temperature
ambient to 325
0
C or more
13
Source
temperature
up to 350
0
C or more
14
mass library
NIST Mass spectral library, pflegermaurer weber library, wiley, pesticide
15
User-friendly
mass software
should be windows based with validation facility
General Condition
2-year warranty of the full system including computer, printer, UPS etc should be provided
Software- suitable for GC along with kinetic software, performance validation software, multi-
component analysis software
consisting of windows based software with features such as multilevel calibration, baseline
correction, all methods of analysis, time programming, plotting and graphical representation of
chromatography
software should be able to control the GC-MS system and date acquisition
the latest configuration of the computer with 2
nd
gen intel core I 3-2120 ( 3.30ghz 1333, 3 MB,
2C), genuine windows ( R) 7 Home basic SP1 64 bit, 4 GB3 DDR3 SD RAM ( 1600 Mhz), 3.5 ―
1 TB 7200 RPM SATA Hard drive ( 1 X 500 GB), DVD Combo drive, 21 ― colour monitor (
TFT), optical mouse, keyboard, HP desk jet/ inkjet colour printer
d. the original equipment supplier/manufacturer should supply the total computer data station
preloaded in the factory
Comprehensive and non- comprehensive AMC rate must be quoted for 5 years from the post-
warranty period
AMC condition ( for both comprehensive and non-comprehensive)
Preventive maintenance- 4 visit per year every quarter
breakdown visit as & when required
A tool kit must be provided free of cost
The total validation package for system and software must be quoted
ISO 9001 certification and GLP compliance.
Training- on-site training for 5 scientists should be provided for at least 10 days of operational
training and 5 days of application training.
Page 29 of 95
Annexure-4
Fourier Transform Infrared Spectrometer (FT-IR)
General Specifications:-
S No.
Specification
1.
Optics
All optics should be self-aligned for easy replacement by the user.
Suitable high stability interferometer with a minimum 5 years warranty.
2.
Spectral range
Spectral range 375-7500 cm-1
3.
Spectral resolution
0.5 cm-1 or better (Non-apodized)
4.
Signal to noise ration
30000:1 or more PP for 1 min. Scan
5.
Fast scan
Minimum 30 spectra/sec or more @16cm-1
6.
Detector
DLaTGS /DTGS in KBr windows
7.
Source
A mid-IR source with a 5-year warranty should be provided. The white
light source should be provided. The white light source should be
provided.
8.
Beam splitter
KBr beam splitter or suitable beam splitter to cover the spectral range.
9.
Wave number
accuracy
0.02 cm-1 or better.
10.
KBr pellet press-hydraulic pellet press capable of making KBr/NaI pellets with controlled
pressure
11.
The system should have the facility for liquid sample analysis (micro-level) using ZnSe and
CaF2 microcell.
12.
The firm should offer micro ATR analytical facility.
13.
Should offer Auto validation facility with the NIST/A5TN traceable validation kit.
14.
Software
The latest version of windows based IR software with the licence to
control all functions of the offered FT-IR system like the facility for
self-diagnostic, software-controlled auto tuning, auto alignment, auto
component recognition and auto-optimization data collection, data
processing, live data display, spectral quality check, spectral search from
the commercial library, creation of the user-generated library, automatic
atmospheric correction, spectral interpretation, GLP compliance, QC.
15.
FTIR Library
i. Georgian crime library
ii. ii. Canadian Forensic Library
16.
Spares/Consumables
The manufacturer/supplier should certify that they will provide all the
spares and consumables at least seven years from the date of installation
of the system.
Page 30 of 95
17.
The firm should offer FTIR Installation kit, validation kit, spares parts for smooth running the
system with necessary certification wherever applicable.
18.
Data station
The latest version of branded PC like
DELL/HP/COMPAQ/LENOVO/IBM etc. Along with the
licence of the original operating system and preloaded software
with 21 LED monitor and laserjet printer.
19.
Two-year warranty of the full system should be provided.
20
The system/manufacturer should be ISO/GLP/GMP/CE complied.
21.
On-site operation and application training should be provided for 05 days for scientists.
22.
Comprehensive and non-comprehensive AMC rate must be quoted for 05 years from the
post-warranty date. AMC condition includes Preventive maintenance-4 visits per year every
quarter and breakdown visit as & when required.
23.
The firm should supply tool kit, hard and soft copies of operation manual and other relevant
literature software and other items wherever possible.
Page 31 of 95
Annexure 5
High- Performance Liquid Chromatograph
General Specifications:-
SNo.
Technical specification
1.
PUMP
High-pressure Binary gradient pump- 02 Nos or Quaternary gradient pump 01 unit
Flow rate 0.01ml/min or lower to 2.00 mI/min or higher
Flow rate accuracy with +/-1% or less
Flow rate precision RSD less than 0.3 % or better
Pressure 6000 psi or better
Degassing unit( Inbuilt / separate)
2.
INJECTOR
Rheodyne type injector or it‘s equivalent and loops 5 microlitres, 10 microlitres & 20 microlitres
with 4 numbers of HPLC syringe (2 nos. of 10 microlitre syringe, 2 nos. of 20 microlitre
syringe)
3.
DETECTOR
Detector-Photodiode array detector- 1 no
Wavelength range 190-700 nm or higher
Wavelength accuracy +/-1nm or better
Spectral resolution -1.2 nm or better
Absorption range/ Linearity up to 2.0 AUFS or better
Photodiode 512 or 1024 numbers
4.
COLUMN
The column having the stationery phase C18 or its equivalent-2 Nos, Column having the
stationery phase C8 or its equivalent-1 Nos or any other column suitable for analysis of
Narcotic drugs and psychotropic substances (NDPS). The quoted columns should be capable of
31 approx.31g NDPS substances with suitable guard column or any other mechanism adopted
to reduce the column contamination.
5.
SOFTWARE
The firm should offer the latest version of windows based licensed software for total control of
the offered HPLC system having data acquisition facility along with performance validation,
peak purity determination, quantitative analysis, calibration and method development etc. The
firm should quote a licensed operating system.
6.
DATA STATION
The principal should supply the required branded (DELL/HP/IBM etc) computer system along
with licenced operating system with 21‖ LED monitor and laserjet printer. The original
equipment supplier/manufacturer should supply the computer data station preloaded in the
factory.
7.
GENERAL CONDITION
The manufacturer and or supplier will provide a 2-year warranty from the date of installation of
the full HPLC system along with the supplied computer, printer and other accessories for self-
sustaining of the entire system.
Comprehensive and non-comprehensive AMC rate must be quoted for 5 years from the post-
warranty date payable in half-yearly post servicing period.
Preventive maintenance- 4 visit per year every quarter and breakdown visit as & when
required. The supplier should also certify that trained service engineers should be available to
attend the emergency within 48 hours.
The manufacturer/ supplier should certify that they will provide all the spares and consumables
at least for 7 years from the date of installation of the system. The supplier should also certify
that they will upgrade the software free of cost whenever such an upgrade occurs.
The firm should supply tool kit, hard and softcopies of operation manual, and other relevant
literature, software and other items where ever applicable.
Training-on site operational and application training for user scientists should be provided for at
least 5 days.
Page 32 of 95
Annexure-6
Ultra-High- Performance Liquid Chromatograph
General Specifications:-
S No.
Technical specification
1.
SYSTEM
UHPLC System must support the pressure of 700 bar (10,000 psi) or better
2.
PUMP
The pressure of 700 bar (10,000 psi) to support <2um particle-sized column.
Flow range of 0.001/min to 10 mL/min.
The Pump should be low-pressure Quaternary Pump with serial dual piston mechanism of
pumping.
The pump has a mixing volume as low as 400μL
The pulsation must be below 0.1% or 0.2 Mpa (whichever is greater)
The pump must have independent compressibility irrespective of mobile phase composition.
Flow accuracy - ± 0.1%or better
Flow precision- below 0.05% RSD or 0.01 min SD (whichever is greater) or better
Gradient accuracy - ± 0.5% of full scale or better
Gradient precision below 0.15% SD or better
The pump must have built-in safety features like Leak detection and safe leak handling, excess
pressure monitoring.
3.
AUTOSAMPLER
The Sampler should support the pressure of 700 bar (10,000 psi) or better
Sample Compartment Temperature Range must be from 440°C
pH Range of the system must be from 1-12 including components like Loop, Injector Valve
and injection wetted parts. Injection range: 0.01–100 μL, min. step = 0.01 μL
Carryover of the Auto sampler must be <0.0004% with a Caffeine solution.
The sample capacity with 1.5ml vials is 216 numbers of vials.
Injection cycle time should be less than 8 sec in specified conditions.
The Auto sampler has to support injection volumes range from 0.01–100 μL. With options to
upgrade injection volume range up to 1mL with multi-draw option.
Injection volume accuracy typically ±0.5% or better
The injection precision must be typically below <0.25% area RSD or better.
The injection accuracy must be typically ± 1% (injection volume 10 μL and water)
The minimum sample volume in vial or well plate required for injection can be as small as 2
μL
4.
THERMOSTATTED COLUMN COMPARTMENT
The column compartment must provide Dual Thermostatting Modes; Still air for highest
efficiency at UHPLC conditions and forced air for easiest method transfer.
Column compartment must have a temperature range from 5°C to85°C. or better
The column compartment must support columns with 300 mm length
5.
MULTI WAVELENGTH PDA
DETECTOR
The wavelength range of the detector must range from 190 to 800 nm
The drift of the detector must be below 1mAU/hat 254 nm or better
No of Photodiodes must be 1024 for better spectral resolutions 0.6nm spectral Bandwidth or
better
The detector must provide a data collection rate of up to 125 Hz or more with spectra
acquisition.
The detector must be able to record 8 channels simultaneously.
Linearity : <5% at 2.2 AU (typically <5% at 2.7 AU) or better
Page 33 of 95
Standard analytical flow cell should be of 10mm path length capable of performing self-
illumination.
The detector must have an internal wavelength calibration using the D-alpha line of the
deuterium lamp
6.
FLUORESCENCE DETECTOR
Two monochromators with concave holographic gratings and elliptic mirrors
The light source must be Xenon lamp with tunable pulse frequency
Wavelength Range for Ex: 200-630nm for Ex: 265-650 nm.
Data Collection Rate must be 100Hz for single-channel acquisition.
Single Spectrum Scans or FL Field Acquisitions: Excitation, emission or synchronous mode.
Spectral Bandwidth of the detector should be (FWHM) Excitation: 20 nm; Emission: 20 nm.
Sensitivity Raman Criteria must be S/N: >2100 or better
Flow Cell Thermostatting 15 °C above ambient to 50 °C
Wavelength repeatability:±0.2nm or better
7.
COLUMN-
Suitable C30, 5µm, ANALYTICAL 4.6 x 250mm column or better with guard column -1Set.
Suitable C18,5um 4.6×250mm analytical column or better with guard column -1Set
Suitable C8,5 um 4.6×250mm analytical column or better with guard column-1set. The
quoted columns should be capable of 33 approx.33g NDPS substances with suitable guard
column or any other mechanism adopted to reduce the column contamination.
8.
SOFTWARE
License version of Chromatography data system software. The software should be a strictly
validated/original licensed copy software with the specific part number mentioned in the
offer [No Pirated version of the software will be allowed and if found the offer will be strictly
rejected]. It should be 21CFR Part 11 compliance. The software should have integrated
database Software should be capable of performing gradient optimization programming 1-9
gradient programming (linear, step, convex, concave) etc. Audit trial should automatically
monitor users action and records a modification history.
9.
DATA STATION-
The principal should supply the required branded (DELL/HP/IBM etc) computer system
along with a licensed operating system with a 21 LED monitor and laserjet printer. The
original equipment supplier/manufacturer should supply the computer data station preloaded
in the factory.
10.
GENERAL CONDITION-
i. The manufacturer and or supplier will provide a 2-year warranty from the date of
installation of the full UHPLC system along with the supplied computer, printer and other
accessories for self- sustaining of the entire system.
ii. Comprehensive and non-comprehensive AMC rate must be quoted for 5 years from the
post-warranty date payable in half-yearly post servicing period.
iii. Preventive maintenance- 4 visit per year every quarter and breakdown visit as & when
required. The supplier should also certify that trained service engineers be available to attend
the emergency within 48 hours.
iv. The manufacturer/ supplier should certify that they will provide all the spares and
consumables at least for 7 years from the date of installation of the system. The supplier
should also certify that they will upgrade the software free of cost whenever such
upgradation occurs
v. The firm should supply tool kit, hard and softcopies of operation manual, and other
relevant literature, software and other items where ever applicable.
vi. Training-on site operational and application training for user scientists should be provided
for at least 5 days.
Page 34 of 95
Annexure-7
Gas Chromatograph (GC) with Head-Space
General Specifications:-
SNo
Specifications
1.
GC
The GC must feature an external color touch screen and
should be capable of installing multiple detectors & injectors on
the basis of the requirement (User-replaceable within two
minutes without the need of service engineer). It should be
possible to control the system completely through a data station
consisting of a computer and GC software and upgradable to
Single Quad MS, Triple Quad MS/MS in future
2.
Oven Temperature
Large Column Oven with a temperature range from 40
o
C
above Ambient to 450
0
C, Temperature Accuracy: ±1%, Heat-
up time from 50°C to 450°C within 4 minutes. Cool-down time
from 450°C to 50°C in less than 4 minutes. The maximum
heating rate of 125°C/min. A number of Ramps /plateaus:
25/26 or More.
3.
Injector temperature
400
o
C orbetter
4.
Injector
split/spitless -Two
Nos with an adaptor
to mount both
capillary & Packed
Columns
The injector should be capable to operate with a capillary, wide
bore and packed columns. The injector needs to permit large
volume splitless injection (up to 50 microliters) without
requiring pressure pulse to quantitatively recover the whole
sample, and without any further hardware requirement.
The injector to allow timed closure/opening of the purge line.
Maximum Temperature: 400 °C
Split Ratio: up to 12500:1
Pressure Range: 0-1000 kPa (0-145 PSI)
Total Flow Setting: Control of split flow in 1 ml/min from 0 to
1:250 ml/min
Pressure range: 01000 kPa (0–145 psi) • Modes: Constant and
programmed pressures and flows with gas saver and septum
purge
Total flow setting: Control of split flow in 0.1 ml/min
increments;
5.
Flow controller for
carrier and carrier
and detector
Automatic digital flow and a pressure
controller/EPC/EFC/IEC must be available
Up to 18 channels of integrated electronic gas control
Pressure set points minimum increments: 0.01 kPa-0.001 psi in
all ranges
6.
Carrier gas
The system should have the facility to use helium & Nitrogen as
a carrier gas in GC system & provision to use Hydrogen as
carrier Gas with adequate safety measure of arresting Hydrogen
Leak.
7.
Safety features
Auto leak detection of carrier gas and automatic shutdown of
GC in case of the leak
8.
GC system should have a gas saver mode
9.
Detector
FID Detector: 02Nos
Sensitivity > 0.03 Coulombs/gC;
Linear dynamic range = >107 (± 10%);
Max Temperature range: 450 °C
10.
Headspace Auto
sampler
12 Vials upgradable to 120-vial capacity with vial loader and 3
removable 40-vial racks, further extendable to 240-vial capacity
in future if needed.
Vial size 10 ml, 20 ml and 22 ml headspace vials with:
Page 35 of 95
Magnetic crimp or screw caps, flat or rounded bottom without
any need of Vial adapter
Oven capacity: Air ventilated oven with 12-seat electrically driven
carousel
11.
Liquid Auto
sampler/Injector
Vial Capacity: 8 vials or more & upgradeable to 150 vials or better
Vial capacity: 2 ml screw top cap
Syringe: 10 μL
Injection Parameters: Max. volume: 5 μL, Mini. Volume: 0.01 μL,
Increments 0.01 μL steps
Syringe Rinsing: Pre and/or Post injection
Vials & Caps for 1000 sample analysis
12.
Validation system
Total validation should be quoted for system and software IQ, OQ
or PQ
13.
Column
DB-1 MS or its equivalent 30 mX 0.32 mm (I.D) x0.25 micrometer
film thickness- one number DB-5 MS or equivalent 30 m X0.25
mm(1.D)x0.25 micrometer film thickness-one number.
3o/o SE -30, OV-17 , Porapak g (1 each)
Trace GOLD TG-ALC1 30Mx0.32Mx1.8 μm or Equivalent -1No
Trace GOLD TG-ALC2 30Mx0.32Mx1.2 μm or Equivalent -1 No (
Pre-column: 5Mx 0.32mm -2 Nos
Micro fluidic 3-port connector for leak-proof dual detector
connection
14.
Software
(with License)
Should be windows-based operating system with validation facility.
The offered software should be suitable for the quoted model GC
along with performance validation.
The facility, multi-component analysis, multilevel calibration,
baseline correction, methods of analysis, time programming,
plotting, graphical representation of the chromatograph and storing
of data
15.
Data station
The principal should supply the branded (DELL/HP/IBM etc)
computer system along with the license of the operating system with
21 branded LED monitor along with LaserJet printer.
16.
All spare parts and accessories for the installation of the full system should be provided along
with one set of the tool kit.
Hydrogen Gas Cylinder with Double Stage Pressure Regulator 01 set
Nitrogen Gas Cylinder with Double Stage Pressure Regulator 01 set
Zero Air Gas Cylinder with Double Stage Pressure Regulator 01 set
Helium Gas Cylinder with Double Stage Pressure Regulator 01 set
Gas Distribution system with Purification Unit for He, H2, N2 & Zero Air
Branded Online 6 KVA UPS with Isolation, SMF Batteries, Battery Rack and Battery
Interconnecting Cables (Three Phase Input &Single-Phase Output) for 60 min. back-up.
Crimper and De-crimper- 1 set
17.
Spares and consumables-supplier/manufacturer should quote and supply essential spares and
consumables including septa (mention item wise) for two years month operation of the quoted
system.
18.
General conditions
i.
The manufacturer and or supplier will provide a 2-year warranty from the date of
installation of the full GC system along with the supplied computer, printer and other
accessories for self-sustaining of the entire system.
ii.
Comprehensive AMC rate must be quoted for 5 years post-warranty date. CMC
should include Preventive maintenance- 4 visits per year every quarter and
breakdown visit as & when required.
iii
.
The manufacturer/supplier should certify that they will provide all the spares and
consumables at least for seven years from the date of installation of the system.
Page 36 of 95
iv
.
Training- The supplier should provide on-site operational/ application training for
laboratory scientists for at least 5 working days.
v.
The firm should supply tool kit, hard and soft copies of operation manual, and other
relevant literature, software and other items where ever applicable.
vi
SPECIAL NOTE:
a) All specifications offered Needs to be supported with original literature as well as
the same literature needs to be available on the website of the manufacturing
company.
b) The offer to be made in details with all technical specification, item details with
part nos (Only line items will not be accepted).
c) All supporting technical literature complying the technical specification needs to
be attached along with the order.
d) Please give reference of at least three users of GC in FSL laboratories in India
with details of the user, email ID & contact nos.
Page 37 of 95
Annexure-8
Accelerated Solvent Extraction System
General Specifications:-
Accelerated Solvent Extraction System complete with pump & fittings which provides operation for a single
sample at a time with the following features & ease of uses in preparing/extracting sample.
Unattended extraction of up to 24 samples.
Samples cell sizes to be available: - 1, 5, 10, 22, 34, 66, and 100 ml.
Collection vial sizes: 60 or 250 ml.
Operating pressure: 1500 psi (100 bar).
Automatic rinsing of the system between sample extractions
Solvent saver mode for further reduction in solvent consumption
Scheduling programming for automated method optimization
It should reduce extraction time and solvent consumption by use of elevated temperature and pressure
during extraction
The requirement of solvent less than 50 mL so to extract a 20gm sample; reducing total solvent usage
It should extract automatically to provide filtered and ready for direct injection or final cleanup sample of
interest.
It should have options for Easy-to-use collection bottles or vials (vials with via tray insert)
It should have convenient multiple-method storage for automatic operation.
It is to come with convenient front panel operation runs methods automatically.
It should have Sensors for temperature, pressure, and solvent and liquid leaks alert the operator to a
problem, sound an audible alarm, and shut down the system if necessary.
It should include solvent bottle, snap-ring pliers, power cords, and gas line fittings with the following
specifications:
Temperature control:
Up to 200 °C; vertical cell orientation with the flow from top to bottom.
Pump: Fluid delivery pressure: 10 Mpa (1500 psi).
Pump flow: 70 mL/minute. Automatic pressure sensor and pressure relief during heat-up.
Fluid Sensors: IR sensors detect fluid level during the collection of extract
Display and Keyboard: Menu operated. LCD 8 × 45 OR similar character display. Method and schedule
editor and storage
Extraction Cells: Seven capacities: 1, 5, 10, 22, 34, 66, and 100 mL cells. Cells feature finger tight cell caps
with compression seal for high-pressure closure
N2 cylinder with Suitable high-pressure regulator & tubing.
Automated sample extraction using flow-through technology with pH-hardened pathways
Automatic extract filtration
Easy-to-fill sample cells with the finger- or hand-tight fittings
Easy-to-use collection vials and bottles
Convenient front panel operation with multiple method storage
Sensors for temperature, pressure, and solvent vapours ensure safe operation at all times
Easy method transfers between systems
3KVA UPS with 60 minutes‘ backup
Page 38 of 95
Annexure -9
Ion Chromatograph
General Specifications:-
1.
PUMP SYSTEM
a)
Binary/Quaternary pumps-02 units (for gradient flow) or any equivalent system suitable for
simultaneous detection of cations & Anions. The system should be suitable for gradient flow.
b)
Flow rate:- 0.01-10ml/min or more.
c)
Reproducibility/Accuracy of flow-<0.1% or better
d)
Flow precision-0.1% or better
e)
Maximum pressure range-not less than 5000 psi
f)
Pump heads and all fluid path should be metal-free (PEEK) or any other good polymer suitable for
analysis of anion and cations.)
g)
Online degasser or equivalent facility
h)
Leak sensor facility
2.
INJECTOR PORT SYSTEM: Dual Rheodyne injector port (Metal-free, PEEK or any other good
polymer ) for simultaneous anions and cations analysis. The firm should offer adequate no of
injection port for simultaneous anions and cations analysis for the offered system.
3.
SUPPRESSOR SYSTEM:
a) Chemical suppressor should eliminate background conductivity & increase the signal to noise ratio
and with continuous regeneration/automatic regeneration or any equivalent technology delivering
the same result. The operation should be automatic and should be able to withstand high
backpressure and high loading or any equivalent suppressor system.
b) The principal /manufacturer will submit an undertaking to supply the suppressor Kit for 7 years
from the date of installation.
4.
DETECTOR (METAL FREE)
Conductivity detector with cell: The firm should offer two (02) nos of detectors for simultaneous
anions and cations analysis for the offered system.
Conductance range: 0-15000microsemens/cm. or better. Block should be thermostated (lower
range-35
0
C or more), Temperature stability -+ 0.01
0
C Auto zero function should be present and
should not control through software. The detector should be suitable to give results up to ppm level
or better.
5.
THE QUOTED SYSTEM(including column and guard column 2 Nos each type and any other
consumables like connectors rings etc.) should be self-sustaining and capable of 38pprox.38g the
following anions and cations simultaneously.
Anions- Nitrate, Nitrite, Fluoride, Chloride, Sulfate, Bromide, Bromate, Phosphate, Perchlorate,
Chlorate, Acetate, Formate
Cations- Lithium, Sodium Potassium, Calcium, Ammonium, Barium.
6.
GENERAL CONDITIONS
a)
The entire offered system for the simultaneous analysis of cations and anions through a single PC
and integrated software. The firm should quote the latest version of the software.
b)
The product should be ISO 9001 certified /CE compliance
c)
Two years warranty of the full system will be provided.
d)
Comprehensive and Non-comprehensive AMC rate must be quoted for 5 years from the post-
warranty period.
AMC condition includes
i.
Preventive Maintenance- 4 visits per year every quarter
ii. Breakdown visit- As & when required.
iii. The total validation package for system and software must be quoted.
7.
DATA STATION
a)
Consisting of windows based software with features such as multilevel calibration, baseline
correction, all methods of analysis, time programming, plotting and graphical representation of
chromatograph.
b)
The software should be able to control the ion chromatography system and data acquisition.
c)
The latest version of branded PC like DELL/HP/COMPAQ/LENOVO/IBM etc. along with the
licence of original operating system and preloaded software with 21‖ LED monitor and laserjet
printer.
d)
The original equipment supplier/manufacturer should supply the total computer/data station
Page 39 of 95
preloaded in the factory.
8.
TRAINING At least 2 weeks of operational and application on-site training must be provided for
5 user scientists.
Page 40 of 95
Annexure-10
Auto DNA Extraction System
General Specifications:-
Electrical Requirements: 220 V, AC 50 Hz.
Open system to accommodate validated kits for DNA extraction from forensic samples.
The system should have the capability to simultaneously process at least 6 samples or more, with
a sample tracking facility.
Built-in system for decontamination.
Cross contamination-free system and provision for monitoring it.
Software for data management and instrument operation/analysis.
Warranty for 2 years from the date of installation and followed by AMC for 3 years.
Page 41 of 95
Annexure -11
PCR Thermal Cycler
General Specifications:-
1. Electrical Connectivity: 220 V, Ac 50Hz.
2. Temperature range 4 ° to 99.9 °C with accuracy +/ or 0.25 °C or better.
3. 96-Well Block for 0.2 mL micro centrifuge reaction tubes or 96 well reaction plate and validated
for use with 4-6 dye-based forensic HID kits.
4. Block format with 6 independent temperature zones capable of standard and fast runs.
5. Ramp rate 3 °C/Sec. or better.
6. Programmed methods for hot-start PCR, cycle sequencing, long PCR, touch down PCR etc.
7. Touch screen colour display of temperature profiles & set up parameters.
8. The system should be licensed for PCR.
9. Validated for forensic DNA analysis as per SWGDAM (Scientific Working Group on DNA
Analysis Methods) guidelines.
10. Warranty 2 years or more from the date of installation followed by AMC for 3 years.
Page 42 of 95
Annexure -12
Real-Time PCR (RTPCR)
General Specifications: -
Electrical Connectivity: 220 V, Ac 50Hz.
96-Well Block Thermal Cycling System with Computer.
Five-color OR more detection capability to perform different forensic applications including
DNA quantitation, SNP (Single Nucleotide Polymorphisms) genotyping, gene expression analysis
and plus/minus assay utilizing Internal Positive Controls.
The specialized optical configuration that supports a broad range of fluorophore dyes.
The system should be supplied with licensed software that design probes and primers.
Sample format 96-well plates and optimized for 20-25 uL. Reactions.
The system should be validated/certified for the latest forensic human/male DNA testing
Quantitation kits.
Software Specifications: -
User-friendly software plate set up, Multi-plate Data Comparisons, Comparative analysis Ct,
Standard Curve, Relative Standard Curve, Allelic Discrimination etc.
Dye discrimination Multi-component algorithm.
Built-in features for sample quality assessment (Human male: female ratio inhibition &
degradation index). Capability for calculating dilution & reaction set up for HID kits.
Warranty 2 years or more from the date of installation followed by AMC for 3 years.
Page 43 of 95
Annexure -13
DNA Sequencer / Genetic Analyzer
General Specifications: -
1. Electrical requirements: 220 V, Ac 50Hz.
2. Polymer-based 8 capillaries automated DNA sequencer upgradable to 24 capillaries.
3. 6 dye-based chemistry.
4. CCD Camera.
5. The latest version of validated GeneMapper ID-X software for fragment sizing (Forensic STR
(Short Tandem Repeat) based human identification) and sequencing with extensive security,
auditing and e-signature capabilities.
6. Genetic Analyzer validated for forensic DNA analysis.
7. Auto sampler plate and facility to use capillaries of different sizes.
8. Matrix Standards.
9. Polymer pouches.
10. Capillaries preferably 36 Cm. size. (2).
11. Anode & Cathode buffers.
12. Deionized Formamide.
13. Forensic STR based kits for HID (Human Identification).
14. Warranty of 2 years from the date of installation and AMC for the subsequent 3 years.
Page 44 of 95
Annexure-14
Comparison Microscope
1. The comparison microscope shall be used for examination and comparison of tool marks and firearm
signatures present on fired cartridge cases/ bullets/ bullet fragments and other forensic physical
exhibits.
2. Basic Comparison Microscope must have following features/provisions:
Mounted on a free-standing working bench/table with motorized Z-movement for height
adjustment of suitable range.
An optical comparison bridge with motorized Z-movement for height adjustment of a
suitable range.
Suitable mechanical stages for the left side and right side preferably with a digital scale or
micrometre in X and Y directions. Coarse and fine focus adjustment for the left and right
specimen stages with independent controls.
Mechanical and/or motorized coaxial drives for X/Y with a range of movement of the
stages of X-50mm/Y-50mm or more and synchronized movement of stages in the X-
direction.
Suitable stage attachments with 3600 rotating basic mounts in horizontal plane and+l1800
angle function with scale divisions.
A tilting binocular viewing tube preferably with variable angle of observation and with
eyepieces l0x/22 FOV or better and with a provision of adjustment of inter-pupillary
distances of the viewers.
The image at eyepiece must be upright and non-reverse and must move in the same direction
as the stage sample being viewed. The images must be colour neutral from left to the right
side.
The width of the image dividing line between the two sample images must be adjustable by
the user and the microscope must have a provision for moving the dividing line for side-by-
side comparisons.
The comparison microscope bridge must offer the following observation modes: side-by-
side view, full left or full right view, and superimposed view.
Provision for simultaneous binocular and video/photo observation/documentation or to
allow full light to be viewed on binocular or full light to divert images to a camera.
Magnification range: from 6x or less to 80x or higher with suitable objectives. Magnified
images must be parfocal and paracentric through the magnification range. Microscope optics
must be apochromatically corrected.
Provision for operational panel supporting all the controls of motorized functions of the
comparison microscope at one place.
Suitable chair with adjustable height, seat and backrest
Dust cover for complete coverage of comparison microscope
Illumination system
An equal and homogeneous illumination of both the samples during examination provided by
one single light source and a splitting system to illuminate the two samples placed or mounted on
two stages shall be preferred.
Suitable light sources including cold light illumination and LED light illumination
The light sources must have all fixtures, holders, spare lamps, fuses, connecting cables, short and
long articulated arms for the fibre-optic light guides, or any other item required for the
illumination systems installation and working.
Holders and Accessories
Suitable holders and accessories for holding or clamping and proper mounting of various shapes
and sizes of 4mm to 25mm calibre cartridge cases in order to examine the firing pin marks,
breech face marks, extractor/ejector marks and chamber marks; and for bullets/ projectiles/
Page 45 of 95
bullet fragments in order to examine rifling/banel marks including suitable bullet/projectile
centring devices/holders.
Holding or mounting surfaces should not alter, erase, add or diminish the firearm signatures on
exhibits. Mounting of exhibits should be slip-proof
5. Digital Imaging Workstation and Software
C-mount, firewire/USB, high definition digital camera for image capture of at least 5- megapixels
image resolution with suitable adapters for optical comparison bridge including licensed software
for image capturing; storing, processing, 45pprox.45g, data transfer, calibrated measurement,
extended annotations, retrieving the image and report generation as well as image database
functions.
The digital imaging workstation with licensed latest windows 64-bit or better operating system
pre-installed, 4GB RAM or more, 1.0T8 SATA HDD storage capacity or more. Licensed office
or equivalent word editor software of the latest version must be installed. The workstation must
come with a USB keyboard, USB optical mouse, 2GB graphics, DVD Read-Write Devices and
a21 or larger high definition LED monitor.
High-resolution color laser printer to print A4 size plain papers and photographic papers for
printing comparison images.
6. Power consumption
AC Power supply of 220t 10% Volts, 50-60Hz
7. Spares and Consumables
The firm must supply spares for a period of at least five (5) years and availability of spares must be
ensured a period of at least for ten (10) years after the date of installation. List of spares and consumables
with cost must be supplied. The firm must provide for a minimum of two (2) years warranty
8. Installation and Training
The firm must do installation and commissioning of the Comparison Microscope at the users site and
provide on-site operators training to four end-users. The firm must have to submit a pre-requisite
requirement for the installation and commissioning of the comparison microscope at the users site after
inspection
9. Operating manuals
The firm must supply the standard operating manuals for the various components of the Comparison
Microscope.
10. Calibration
Comparison microscope must have the facility of calibration for its various measurements.
Page 46 of 95
Annexure -15
SEM-EDXA
General Specifications:
1.
Resolution
At least 1.2nm at 30 kV or 1.5 nm at 15 kV and 3.0 nm at 1 kV)
2.
Magnification
50X to 500,000 X or better
3.
Acceleration Voltage
0.1 to 30 kV (or better )
4.
Chamber
Large chamber with at least 7 accessory ports. Anti-
vibration table must be inbuilt.
5.
Probe Current
1 nA to 1μA or better suitable for all applications.
6.
Detectors
a) High sensitive Everhart-Thornley SE detector
b) In-Lens SEI detector for high-resolution imaging in High
Vacuum at lowkV OR low vacuum/ variable pressure mode.
c) IR-CCD camera
d) High-resolution state of art backscattered
detector(BCD) for using both in high vacuum and
variable pressure.
e) The detector should have the capability of high
sensitivity for low kV analysis.
f) Specify built-in automatic/ manual control for
contrast and brightness.
7.
Electron Gun
Thermionic emission type Tungsten source
8.
User Interface
Keyboard, Mouse, Control Panel with multifunction for the control
and adjustment of frequently used SEM parameters
9.
Electron Optics
Beam deceleration technology or equivalent for high-resolution
imaging at low kV. Ease of operation is desired.
10.
Vacuum System
Fully automated microprocessor-controlled vacuum system
a. Suitable vacuum system equipped with ion pumps, turbo-
molecular pump & rotary pump.
b. Chamber pressure better than 10
-4
Pa / 4 x 10
-6
mbar
c. Pump downtime should be less than 5min
11.
Digital imaging and
processing
It should have the following capabilities:
a. Design of the imaging and processing should be
optimized for SEM-EDX electron microscopy
b. Image Frame Size: Selectable up to pixel density of 4096 x
3536 or better
c. Image post-processing
d. 4 detector inputs and signal mixing or above, extendable up to
8 detect or inputs
Image Display
a. 22 high-end TFT flat screen or better
b. Standard data zone includes magnification, working
distance,scale bar and date Custom data zone
c. Multiple point-to-point and line width measurement
systems freely adjustable for orientation
d. Line profile display
e. Images can be viewed live, averaged or integrated
Image Storage
a. 1 TB hard disk or better
b. Front-panel USB ports. CD/DVD recorder
c. Storage of SEM images on the hard disk in standard TIFF,
BMP, or JPEG Formats and in 8-bit or 16-bit depth
Page 47 of 95
d. Operating conditions easily stored and file management
through Microsoft® Windows operating system
12.
Essential Accessories
a. Chiller.
b. Compressor.
c. Interface among SEM EDX/EDS.
d. 3 Spare Filaments
e. 20 number of single stubs and 10 number of multiple
sample holders
13.
Integrated EDX/EDS
EDS
Integrated EDX/EDS system
a. LN2 Free SDD detector with 30 mm
2
crystal area and 130 eV
resolution or better.
b. The elements detection range should be from Beryllium (Be) to
Uranium(U).
c. The EDX/EDS should be capable of selective element
mapping, line scan, selected area analysis, quantitative analysis,
qualitative analysis, multi point analysis.
d. Supplied EDX/EDS server & analysis software should be
capable of performing data acquisition, storing and transfer in
common Windows-based application format, qualitative &
quantitative analysis, line scanning, elemental or dot-mapping
(area) including spectrum imaging and phase mapping with
specimen drift correction.
e. Thin-film analysis software with nanometer-scale resolution in
both space and depth capabilities should be quoted.
15.
Warranty,
Training and Support
a. Three years of comprehensive warranty
b. Necessary on-site training must be provided.
Page 48 of 95
Annexure -16
Ballistics Data Management System
General Specifications:
VELOCITY MEASUREMENT SYSTEM
The system should have the provision to measure any velocity in the ang of 10 2000 m/s (or more)
with an accuracy of up to 0.01%. The system should also have the provision of measuring velocities at a
minimum of three points of the trajectory simultaneously so as to evaluate the muzzle velocity, striking
velocity and remaining velocity. The non-contact velocity measurement equipment will consist of optical
detectors, light sources, multi-channel timer units, workstations and various connectors, cables etc. as
described in the subsequent section. The calibration protocol of these measurement systems should also
be provided and suitable training is given during the pre-dispatch inspection stage.
1.1: OPTICAL DETECTOR 06 Nos. (3 sets)
Projectile calibre: 4mm to 10mm or more
Projectile Velocity: up to 1500 m/s
Sensitivity: The optical detectors should be able to work in ambient light and respond to
less than 0.1% change in ambient light level.
Operating temperature: up to 60º C
Provision of automatic/fixed alignment with the light source
Mountable on rails with brake/locking system
Apparatus must be sealed against moisture.
Commissioning, testing and training at a site
1.2: LIGHT SOURCE 06 Nos.
Coupled with optical detectors
Rugged construction
Power supply: 200 260v AC ±10%, 50/60 Hz
Humidity: up to 95%
Operating temperature: up to 60ºC
Mountable on rails with brake/locking system
Commissioning, testing and training at the site
1.3: MULTI-CHANNEL COUNTER 1Nos.
Each multi-channel counter will support optical detectors (minimum 3 sets)
Storage capacity: ≥ 256 velocities or rate of fire measurements
Rate of fire: 1000 rounds per minute or more
Operating temperature: up to 60ºC
Humidity: up to 95%.
Sufficient cable for minimum 100 m installation.
1.4: WORKSTATION (1 Nos.) AND ACCESSORIES
Workstations comprising of latest and high-end processors along with hardware
accessories like high-end graphics, data writers (DVD/CD), wireless keyboards, mouse,
external memory devices, Ethernet modules, high-end printers etc. The mainframes
should be interconnected.
OS and other s/w: High-end graphics s/w, Windows (latest), data acquisition s/w, range
control, office suites, networking.
Commissioning and testing at the site.
Page 49 of 95
1.5: UNIVERSAL BREECH SYSTEM AND STANDARD BARRELS (1 EACH)
One universal receiver for above barrels, recoil rest with barrel support and laser
alignment tool.
Remote firing units and security equipment for the above.
Cleaning and spares kit.
Mountable on rails with brake/locking system
Commissioning, testing and training at the site
1.6: STANDARD BARRELS
One test barrel (rifled) as per ANSI/SAAMI standard for 5.56x45 mm ammunition.
One test barrel (rifled) as per ANSI/SAAMI standard for 7.62x51 mm ammunition.
One test barrel (rifled) as per ANSI/SAAMI standard for 7.62 x 54 mm ammunition.
One test barrel (rifled) as per ANSI/SAAMI standard for 7.62X39 mm ammunition.
One test barrel (rifled) as per ANSI/SAAMI standard for 9x19 mm ammunition.
Cleaning and spares kit
1.7: FIREARM REST TABLE (1 No.)
For mounting the different calibre firearms and withstanding the jerk caused during
firing.
Height 1200 ±100 mm, Weight less than 500 KG.
PLC controlled auto trigger system with touch panel.
Dual vice model for mounting shoulder arms and handguns.
Barrel mounting vice and butt rest to avoid any recoil during firing.
Lockable anti-vibration pads to absorb recoil.
Arrangement for tilting the firearm.
Caster wheel and toe provision with locking system.
Cabinet for storage of guns/tools.
1.8: TARGET TABLE (1 No.)
Target table in X, Y, Z and C axis with the remote control system
Height of table platform 1200mm ±300mm movement.
Z-axis movement ±300mm
Stainless steel dial plat with diameter 1000 ±100mm rotating 360 degrees with step
movement of 5 degrees each.
Caster at the base for movement and adjustable mounting foot for firm resting.
1.9: HEAD FORM (2 Nos.)
Head Form (aluminium) for mounting large, medium and small-sized BP helmets as described in
NIJ standard.
1.10: TROLLEY (1 No.)
Manually driven hydraulic trolley of 400 kg capacity.
Fork span wit in the range of 600 ±50mm, Fork length 1150±50mm and height of
fork adjustable from min. 90mm to the max. 1500mm.
1.11: BULLET TRAP (1 No.)
For stopping the bullet with velocity 4500 f/s.
The size will be 3m × 3m (vertical blinds).
Commissioning, testing and training at the site.
Page 50 of 95
Annexure -17
X-Ray Fluorescence Spectrometer
General Specifications:-
S. No.
Technical
Parameter
Specifications
01
H.V. Generator
H V generator should be 3 KW or more, 60 KV (with 1 kW
steps), 100 mA (with 1 mA steps), Stability of the Generator
should be ± 0.005% (for both kV and mA, against ± 10%
input variation) or better.
02
Excitation
Source/probe beam
Rhodium (Rh)/ Tungsten (W) anode based X-ray tube (about
4 kW) capable of operating up to 40 kV, suitable for analysis
of low Z and high Z elements. Appropriate filters for
detection of various elements with a wide difference in atomic
numbers.
03
Optics
Optics should be best suitable for analyzing powder and
solids samples. The system should have proper dust trap for
protection of the optical chamber.
04
Detector
High Resolution & large area Fast Silicon Drift Detector(Fast
SDD) with the capability to acquire X-ray data/count rate
greater than 100 kcps (kilo counts per second).
05
Operating
environment
Temperature: -5 °C to +50 °C or more; Humidity: 10 % to 90
% relative humidity non- condensing.
06
Resolution
<145 eVFWHM @ 5.95 keV Mn K
α
line.
07
Elemental Range
F(9) to U (92) OR Mg(12) to U (92)
08
Sample Chamber
and Automatic
Sample Changer
The system should be offered with at least 12 positions
Automatic Sample Changer. Sample chamber should be easily
accessible for the user for day to day cleaning purpose.
09
Vacuum Pump
Single vacuum pump with dust trap should be offered
10
Spectrometer
Chamber
Should have a system for Thermal stabilization preferably
heater system to keep optical chamber temperature steady.
11
Data acquisition
Data acquisition and user-friendly instrument operation and
data processing and file storage software and hardware
(PC/Monitor/Printer/Licensed software) should be offered.
12
Measurement area
Minimum spot size shall be 5 mm or less
13
Safety provisions for
the instrument and
operation
X-ray enabling only in the presence of the sample using an IR
proximity sensor. Safety provision for X-ray detector.
Shock and water-resistant case.
14
Data storage and
transfer
Internal memory: 500 GB or better
Data transfer : Through USB port for connection to a
computer for data transfer. The instrument software should
support Microsoft windows 8/10 or higher version of the PC
for data transfer and readable through Microsoft-
Word/Excel.
15
Operating
Environment
/Power supply
Line Voltage: 230V±10%, 50/60Hz ± 5% (without any step-
up or step-down transformer) with protective RF ground.
Appropriate DC power supply cord with an adapter to
operate from an AC power source (100-240 VAC, 50-60 Hz.)
shall be provided.
Page 51 of 95
16
Device/
instrument software
The operating system of the instrument shall be licensed and
Microsoft Windows-based software.
There should be a backup of all essential software related to
OS and application software in the form of the flash drive.
Spectrometer software should be compatible with Window
8/10, 64bit, including multipoint background correction, line
overlap correction, ratio calculation, calibration curve,
qualitative analysis incorporating matrix correction,
recalibration program, data storage & retrieval facilities.
17
Application
software for
composition analysis
Built-in calibration check at the start including standard
specimens.
Automatic grade identification using a database of standard
grade library (≥300 grades) covering various International
standards.
Standard Library:- It shall include grade library.
18
Training
Hands-on training including details of hardware components,
OS, application software, calibration checking, chemical
composition analysis of various standard specimens and
others, troubleshooting, report generation, data transfer and
saving the spectra and composition results to other computers
through USB.
19
Certificates/
Manuals
ISO Certificates.
Full details of authorized service partner in India for
providing after-sales technical support.
20
Consumable
Accessories
As per requirements of the instrument, consumables &
accessories should be supplied along with equipment for the
smooth functioning of the XRF Analyzer.
21
Warranty
3 years
22
System Accessories
Suitable UPS ( 15 KVA, 30 min back up) to run the system
and any other recommended items or support system.
Suitable branded computer (1 no) with the printer,
Suitable cooling system (water chiller1 no).
Sample holders.
Page 52 of 95
Annexure -18
X-Ray Diffractometer
General Specifications:-
I.
X-RAYGENERATOR
Power
The maximum continuous output power of 3 KW or higher
Input voltage range
208 240 V
Rated Voltage
20-60 KV or better range (in steps of 1 KV).
Rated Current
2-60 mA or better range (in steps of 1 mA).
Stability
0.01% per 10% main variations
Cooling
The system should have a requisite cooling system external
chiller/heat exchanger etc. for smooth and stable running.
Control
Fully controlled through Windows-based PC software
Safety Device
Abnormal cooling water, flow rate, water Pressure, Temperature
detection, abnormal XG load (overload, line current, abnormal low and
high voltage, emergency stop switch, leak breaker), shutter malfunction
detection
II.
X-RAYTUBE
Insulation
Ceramic
Anode material
i.
A Cu anode; The system should also be able to work with other
X-ray sources like Mo, Co, Fe, or Cr.
ii.
A Co anode must also be coated.
Maximum power
2 kW or higher
Focus
The fine focus, long line fine focus with Ni K
Filter. Facility to
switch between line and point focus applications without
disconnecting cables and utility lines or unscrewing the X-ray tube.
Tube Shield
Electromagnetic shutter interlocked with radiation enclosure
III.
GONIOMETER
Geometry
Vertical (keeps sample horizontal and stationary)
Operation mode
Theta Theta
Scanning mode
θ
D
S
independent or coupled
2 theta
measurement range
-90 to 165 deg or better range (with the capability of
transmission mode)
Scan Speed
Between 0.001 to 1.25 deg/sec or better
Slew speed
9 deg/sec or higher
Min step size
0.0001
o
or better.
Diameter
The minimum diameter of 480 mm; without major alignment
procedure will be preferred.
Page 53 of 95
Data quality guarantee
The manufacturer must submit data quality guarantee certificate
from the factory on the angular position (<= 0.01 deg 2theta over
the entire angular range) to be carried out on NIST sample. The
same NIST sample should be in the scope of supply for
evaluation.
IV.
OPTICS
Slit type
-
Automatic Computer Controlled & Programmable Variables lit
-
Alignment-free and tool-free change of optics
-
Should be used either in the fixed or non-fixed mode
Primary / Incident slit
Motorized slit and parabolic multilayer mirror for changing
from Bragg & Brentano focusing optics to parallel multilayer
mirror 0.05-7.00 mm or better; 0.01 mm Step or better
Secondary / Receiving slit
Motorized slit and parabolic multilayer mirror for changing from
Bragg & Brentano focusing optics to parallel multilayer mirror
optics
0.05 20.00 mm or better, 0.01 mm Step or better.
Slit specification
Motorized slit and Soller collimator (minimum 2 suitable sizes)
along with Anti-scatter slit assembly to reduce air scattering
particularly for low angle measurements; for incident beam and
detector
Fluorescence suppression
Suitable hardware and software for suppressing secondary
fluorescence must be included.
Height limiting slit
Necessary height limiting slits are to be included
Optical system alignment
Complete automatic alignment without manual intervention for
the alignment of source height, source angle, mirror optic, crystal
optic, slit height, sample surface and detector angle.
V.
DETECTOR
Type
-
Solid-state high-speed detector with a minimum 150channels
/ semi conductor strips equipped with diffracted beam
monochromator; must be maintenance-free and must work
without using any cooling agent or gas.
-
It should offer good angular resolution and perfect profile shapes
with no defective channels/pixels and operable for all kinds of
radiations ex. Cu, Co, Cr, Fe & Mn.
Efficiency
>98% for Cr, Co, and Cu radiation
Spatial resolution (pitch)
Minimum 75 micro metre
Capture angle
>2.5° 2Theta angular coverage at 500 mm measurement circle
diameter on both X & Y direction and all channels/pixels should
remain active for a minimum 5 year from the start of use. In case
if any of pixel/channel found dead within this period supplier has
to replace with a new detector FREE of cost.
Counts
Minimum 100,000,000 cps or better.
Energy resolution
<680 eV for Cu radiation at 25°C
Fluorescence suppression
Suitable hardware and software for suppressing secondary
fluorescence to be offered.
Page 54 of 95
Low Angle Data
Capable to collect high-quality data starting at angles as low as
0.15° 2Theta.
VI.
SAMPLE STAGES
Sample stage type
-
Computer-controlled rotating sample stage with the ability to control
& vary the rotating speed for orientation studies with suitable motors.
This stage should facilitate both reflection & transmission
measurements with sample horizontal.
-
Capable to accommodate samples of different thickness with
minimum 10 numbers of sample holder (for spinning and non-
spinning)
Sample stage for stress,
Motorized Chi (-3
o
to +90
o
or better) and Phi rotations and Z
texture and thin-film study
translations suitable for common thin film, texture or residual
stress samples and other samples with load capacity minimum
250 g.
Optional item: The basic high-resolution thin film attachment
should include incident Ge (220) channel-cut monochromator to
remove K2doublet.
For small samples
Silicon zero background sample holders to hold small sample size
(150 mg) flat and cavity type or equivalent, 2 Nos each should
be supplied which can be used with a rotating sample stage.
VII.
SAFETY AND RADIATIONENCLOSURE
The XRD system must be fully radiation protected as per the international radiation safety
norms.
The offered XRD system must comply with the requirements of the Machinery Directive
2006/42/EC for European Manufacturer or the equivalent norm for non-European
Manufacturer.
Open & Close Door System (with a Pb-contained Acrylic Resin Windows)
Lead equivalent: 0.5mm Pb equivalent.
Warning Light: Installed independently on the top of the Radiation Enclosure.
VIII.
SOFTWARE
The PC based software must be capable of both qualitative and semi-quantitative analysis; and
easy simulation and automatic refinement / smoothening of measured data.
The system should have provision for interfacing with computer and data analysis software
shouldbecapableofsimultaneousdatacollectionandcarryouttheanalysiswitha minimum of five
multi-user protected licenses for independent systems. The software should have a facility for
remote operation and diagnostics of the instrument.
Highly integrated software from XRD manufacturer employing various profile fitting techniques
for working on below applications should be offered under options.
Single Line Fitting up to Whole Powder Pattern Fitting.
Whole Powder Pattern Decomposition
Structure determination with features like Fourier series, Charge Flipping etc. and refinement by
features like Rietveld method, Two-Step Method, Rigid body, preferred orientation corrections
etc.
The software should work with Fundamental Parameters methods line profile, shape analysis
by fundamental parameters approach (FPA) or equivalent.
Quantitative phase analysis with various whole pattern methods for amorphous as well as
crystalline phases by employing Rietveld, Internal Standard, External Standard, phases with
known or unknown structure. Measurement and data collection with variable counting time
feature should also be a part of the software.
Page 55 of 95
The diffraction data may be capable of peak search, k-alpha1 and k-alpha2 separation, integrated
intensity calculation, multiple recording of raw data files, measurement condition display and
editing, data conversion to ASCII and general text format output as colour-coded diffraction
profiles, reflection lists, or as raw data files, or in ASCII/ CSV forms, and exportable to popular
platforms like MS Excel. The offered data acquisition software should run on Windows
platform format for additional user manipulation.
The software should be enabled with the latest Rietveld algorithm for standard-less quantitative
analysis.
The manufacturer must offer their licensed software developed by them with certificates along
with media and exhaustive operating manual(s). Licenses for all the databases and computational
software should be for a minimum of five years with updates. Periodic updates of all software
should be provided free of cost for a period of a minimum of five years.
Latest ICDD Database with Minimum five User 5 years License must be offered.
The application software package for residual stress and texture (pole figure) and orientation
distribution function (ODF) must be included with a suitable application-oriented attachment
for residual stress and texture measurement.
Page 56 of 95
Annexure -19
Atomic Absorption Spectrometer
General Specifications:-
S No
Technical
Parameter
Specifications
1
Atomizer System
1. The system should have software-controlled automatic vertical
and horizontal alignment of the flame burner head for
optimum light transmission.
2. The system should come with standard 10 cm titanium burner
head for Air-Acetylene flame for better absorbance.
3. The system should have software-controlled flame ignition.
4. A suitable air compressor should be available with the system.
5. The system shall have a possibility for direct solid sample
analysis or alternate mechanism for 56 approx.56g solid
samples.
2
Light Source/ Lamp
support
1. The system should have a minimum 4 to 6 lamp holder
including Tungsten, halogen, hollow cathode lamps with long
life and emission intensity at ≥3000 ºC, or equivalent.
2. The system should have a provision of automatic lamp
selection, slit width selection, and wavelength selection.
3. The light source should be effective across many wavelengths
and stable over its lifetime.
4. The system should have a built-in power supply for all types
of lamps.
3
Sample Introduction
System
1. A high sensitivity nebulizer system including impact-based
and flow spoiler with corrosion resistant against the acids.
2. Corrosion-resistant spray chamber.
4
Optical system/
Monochromator
1. A true double beam spectrometer system with high light
throughput.
2. Monochromator system with diffraction grating ruling density
with at least 1800 lines/mm blazed in both the UV and Visible
regions.
3. A focal length of minimum 250 mm and a reciprocal linear
dispersion of 1.6 nm/mm.
4. Variable slit width between 0.2 to 2.0 nm with automatic slit
selection.
5. The system should have the maximum light transmission for
the best detection limit with least maintenance and with
updated technology like fibre optics for transmission of light.
6. The system should automatically adjust to changes in lamp
intensity for stable baselines and compensates for drift multiple
times per second.
7. The system should have a fast start-up and exceptional long
term stability without recalibration.
8. 4-step bandwidth selection with automatic switching 0.2, 0.7,
1.3, 2.0
5
Detector
The system should have P.M.T Detector with high quantum
efficiency. The system should have operator selectable read time
from 0.1 to 120 sec.
6
Wavelength
185 900 nm or better.
7
Background
Correction system
High-speed continuum source (Deuterium lamp) for Background
correction
Page 57 of 95
8
Gas flow system
1. Software controlled flame ignition and automatic changeover
of oxidant.
2. Fully software controlled oxidant and fuel gas flow monitoring.
3. All safety interlock built-in and additional features like burner
head interlock, nebulizer End cap interlock and drain interlock
to be built-in.
9
Sensitivity
> 0.9 absorbance with the precision of <0.5% RSD from 5
second integration for 5 ppm Cu standard for flame operation.
10
Accessories
1. Vapour/Hydride accessory to connect with flame.
2. HCL lamp for suitable elements like Mn, Fe, Ca etc., and high-
intensity electrode less discharge lamp for Hg, Se, Pb and As
elements. Built-in suitable power supply for these lamps.
3. Individual element standards 1000 ppm.
4. Suitable PC with a colour printer and suitable UPS.
5. Acetylene gas and Argon gas with regulator and purification
panel
6. Pyrolytic coated graphite
7. Vapour generator: Continuous flow type for fast determination
of hydride and cold vapour formers such as Se, As, Hg, Te, Bi,
Sn, Ge at low detection levels
8. Vaporizer to improve detection limits/sensitivity of elements in
the test scope.
9. Syringe, Plunger, Sample vials 1.1ml, Sample vials 2ml,
Capillary assembly, Cuvettes, T-cell for mercury and arsenic
and Gas purifiers.
10. Oil-free Air compressor.
11
Warranty
3 years
12
Voltage stabilizer
15 KVA Voltage stabilizer
13
Power supply
Shall be suitable for operation on 220 V-240V and 50/60 Hz mains
power
14
Safety features
The system should include are-circulating water
chiller/cooler(below0°Cand>3Lofwater; compatible with 220 V-
240 V power requirement) also includes a safety system capable of
controlling the cooling water (in terms of heat and pressure), gas
pressure, mains power, furnace block and over current.
15
Training
Hands-on training through an authorized representative shall be
provided to users at purchasers site for the safe operation of the
Instrument including details of hardware components, OS,
application software, calibration checking, analysis of various
standard specimens and others, troubleshooting, report generation,
data transfer and saving the spectra and composition results to
other computers through USB.
16
Certificates/
Manuals
1. ISO Certification.
2. User‘s manual of the instrument that includes safety
instructions during operation and maintenance of the
instrument.
Page 58 of 95
Annexure -20
Inductively coupled Plasma Atomic Emission Spectrometer
(ICP-AES)
General Specifications:-
1.
PLASMA GENERATOR
a. RF Generator
Solid-state generator with less warm-up/Startup time
b. Frequency
Crystal controlled40MHz/ 27 MHz with stability better than 0.01%
c. Power
Computer adjustable from 800 - 1550 watts for 40MHz/1400-
2000wattsorwiderisbetterfor 27 MHz
d. Controls
Fully computer-controlled with programmable power, start
and shut down of the plasma.
e. Gas flow controls
Computer-controlled mass flow controllers for all channels should be
available for control of the flow rates of Plasma, Auxiliary, Nebulizer gas
f. Plasma mount
Plasma should be vertically mounted & radially viewed
g. Re-circulating
Chiller
The external chiller with details should be provided for an external need
for RF coil cooling,
2.
OPTICAL SYSTEM
a. Wavelength range
165 to 770 nm
b. Grating
Original, ion etched holographic grating/Echelle cross dispersion with
ultra-violet fused silica prism
c. Focal length
More than 0.5 meters for holographic grating/ More than 0.32 meter for
Echelle grating
d. Entrance slit height
Maximum to view complete plasma height in one shot
e. Optical Resolution
Better than 10 pm up to 320 nm OR (7 pm at 200 nm or better)
f. UV purge
Nitrogen/argon purge/gas sealed for 165 to 190 nm.
g. Stabilization
The temperature of the optics should be internally regulated to ≤ 2° C
h. Grating density
More than 50 lines/mm for Echelle grating/more than 2000 lines/mm
for holographic grating.
3.
DETECTION SYSTEM
a. Type
CID/CCD/PMT
b. Dynamic range
Greater than 5 orders of magnitude
c. Cooling
If CCD is used it should be cooled internally with multistage Peltier
d. Range
Spectral range from 165 to 770 nm OR 167 to 853 nm
e. Blooming
The detector shall not be subject to any blooming effect from nearby
wavelengths, or excess concentrations.
4.
SAMPLE INTRODUCTION
a. The kit which
includes a torch,
spray chamber,
nebulizer &
tubings
1. Aqueous Kit
2. High Solid Kit
3. HF Kit
4. For Organic Kit
b. Sample pump
More than 10 rollers peristaltic pump with 3 or 4 channels. Computer-
controlled rinse and flush times, as well as computer-controlled sample
flow rate.
5.
PLASMA TORCH
a. Torch type
Completely demountable and vertical mounted
Page 59 of 95
b. Alignment
It should be possible to remove and replace the torch, spray chamber,
nebulizer etc.
c. Viewing height
Maximum plasma should be viewed at once, to minimize interferences.
6.
COMPUTER SYSTEM AND SOFTWARE
a. Configuration
The configuration of the computer should be suitable to the unit supplied
with a 21 inch full HD monitor. An optical mouse, optical keyboard and
laser printer.
b. Software
1. The software must provide preprogrammed conditions for a variety of
matrices.
2. The software must provide optimization by the element of sensitivity,
speed and precision based on users input.
3. Database availability and provision in the software for matrices,
elements run most often and standards also.
4. The software must allow for both the export of data and import of
sample details in a suitable format.
5. Software shall provide a library for maximum emission lines in all
orders.
7.
INSTRUMENT PERFORMANCE
a. Precision
The in-run precision for a suite of 3 elements Pb, Al, Cd at 1 ppm shall be
less than 1.5% Relative Standard Deviation (RSD).
b. Stability
The ICP spectrometer shall be able to demonstrate an RSD of less than
2% for all elements.
c. Precision and
stability with
dissolved solids
The ®CP must be able to demonstrate running a 20% NaCl solution for
more than 4 hours without clogging, and achieve a precision of less than
5% RSD for 5 ppm Pb solution and be able to get a Limit Of Detection
(LOD) of less than 10 ppb for Pb using the 20% NaCl blank solution.
d. Radial Detection
limits
The instrument shall offer < 10 ppb detection (LOD) at 3 sigma with a
RADIAL ICP.
8.
CONSUMABLES KIT
a.
Should contain:
Torch tubes for HF application, High Solid applications and aqueous
applications.
b.
Tubing for each of the above applications.
9.
AUTO SAMPLER AND HYDRIDE GENERATOR
a.
Random access auto-sampler for unattended analysis, with auto-
shutdown at end of batch analysis for samples.
b.
Hydride generator for an increase in sensitivity for hydride forming
elements.
10.
OTHER ACCESSORIES TO BE SUPPLIED
a.
Suitable fume exhaust system
b.
Swagelok dual-stage regulator.
11.
WARRANTY - Three years
12.
STANDARDS
A set of standard solutions for the system calibration with NIST traceable
certificate should be supplied.
Page 60 of 95
13.
TRAINING
Should be provided after installation. The training should be for a period
of one week covering all aspects of the software.
14.
UPS
Suitable UPS should be provided
Page 61 of 95
Annexure-21
Hardware Based Forensic Imaging Device
General Specifications: -
S.No.
Tool/Facility
Specifications
1.
Hardware Imaging
Device
Write Blockers Kit
Tableau Duplicator
Falcon
WRITE BLOCKERS KIT
The kit should have the blocker devices for the storage media
having the interface of SATA, IDE, Firewire USB2, USB3, e-
SATA, Micro-SATA, Combine SATA and ZIF, and SAS (Serial
Attached SCSI) along with their respective connecting cables.
TABLEAU DUPLICATOR
Image SATA, USB 3.0, and IDE hard disks natively
Image SAS drives with the SAS Expansion Module
Standard operations:
Disk-to-Disk (clone) duplication
Disk-to-File (image) duplication
Format, Wipe, Hash (MD5 or SHA-1)
HPA/DCO detection and removal
Blank Disk Check
FALCON
1. Capable to acquire/clone data from one-to-one, two-to-two
destinations media.
2. Capable to acquire/clone data at the rate of 5GB/Min. (or
more).
3. Capable to acquire/clone data in Drive to Drive, Drive to
Uncompressed segmented files and Drive to Compressed
segmented files format to support analysis in various forensic
analysis tools.
4. Capable to cross copy support to IDE, SATA, eSATA, micro
SATA, SCSI, SAS, ZIF and USB interfaces and combine-SATA
etc.
5. Capable to authenticate the data with at least two hashing
algorithms.
6. Capable to identify and acquire HPA and DCO Areas of the
suspect media.
7. Capable to acquire data directly from the live system/content-
based image.
8. Capable to boot/Mount the suspect media virtually in a write-
protected environment from the preview of live data.
9. Capable to acquire data from suspect RAID (all RAID
Configuration) drives.
10. Capable to acquire data over a network.
11. Capable to generate the log of the processes.
12. Capable to search for the keywords in the suspected media.
13. Preloaded operating software.
Page 62 of 95
Annexure- 22
Software for previewing, Imaging and analysis of digital Media
(Encase software)
General Specifications:-
SNo.
Tool/Facility
Specifications
1.
Software for the
Forensic
Previewing
Imaging and
Analysis of
Digital Media
(Encase
Software)
1. Should support preview of suspect media.
2. Should support acquire data from various types of storage media.
3. Should support acquiring data in multisession.
4. Should support various operating systems viz, Windows, Linux, Unix,
Sun Solaris, Macintosh, Apple‘s iOS, Android OS, Blackberry OS,
HP‘s Palm OS, Nokia Symbian, Windows Mobile OS etc.,
5. Should support automated processing of evidence with the capability
to indexing, auto de-NISTing, file signature analysis, hash analysis,
Protected File signature analysis, hash analysis, Protected file analysis,
Expand compound files. File parser for Emails (PST, NSF, DBX,
EDB, AOL, MBOX), extract internet artifacts, timeline analysis,
System Info Parser, IM Parser (AOL, MSN, Yahoo), File Carver,
Windows Event Log Parser, Windows Artefact Parser, Unix Login,
Linux Syslog Parser, etc.
6. Should support decryption of disks, volumes, folders and file of the
suspected media having various types of encryption.
7. Should support custom programmes for processing of data.
8. Should support various file formats for viewing the data.
9. Should support iOS Physical image analysis.
10. Should Support 64 bit systems.
2.
Software for the
Forensic
Previewing
Image Mounting
Password
Cracking and
Forensic Analysis
of Digital Media
(FTK 6
M/S. Access
Data)
1. Should support windows, Apple, UNIX and Linux operation system
for analysis of the media.
2. Should support RAM analysis.
3. Should Support PLIST, SQLite database, Apple DMG and
DD_DMG disk image and JSON file, etc.
4. Should support Automated processing of evidence with the capability
to indexing, auto de-NISTing, Recover files and partitions, detect
deleted files, file signature analysis, hash analysis, Recover Folders,
Expand compound files, parse Email. Databased, extract internet
artefacts, timeline analysis, system Imfo Parser, IM Parser (AOL,
MSN, Yahoo), File Carver, Windows Event Log Parser, Windows
Artefact Parser, etc.
5. Should support the registry log analysis.
6. Should support auto-detection of password-protected files, decryption
of various types of encrypted files.
7. Should support the generation of CSV based reports for timeline
analysis.
8. Should support automated explicit image identification by skin tone
analysis.
9. Should support automated file indexing and statistical report
generation of each file type.
10. Should Support auto carving of files and adding the carved files to the
file tree structure.
Page 63 of 95
3.
Forensic Tool
Kit for Incident
Response of
Digital Crimes
(VPER Kit)
1. Mobile workstation:
Processor: Intel Core
TM
i7 (3920XM)
Operating System: Windows 7 Ultimate 64-bit or latest version
Memory: Up to 32GB2 DDR3 SDRAM at 1600Mhz or higher
Chipset: Mobile Intel
®
QM77 Express Chipset
Video Card: NVIDIA Quadro K5000M with 4GB GDDR5
dedicated memory
Display: 17.3FHD(1920x1080), Premier Color technology, IPS,
with a view, anti-glare, LED-backlist.
Hard Drive: One TB SATA hard drive.
Optical Drive: DVD-ROM, Slot load DVD+/RW, Blu-ray Disc
writer;
Connectivity:
Wired: Integrated Intel 82579M/V 10/100/1000 Gigabit
Ethernet
Bluetooth: Wireless
Should have the facility for hard drive swapping
2. The Kit should have Following tools for search, seizure and analysis
of digital evidence.
Hardware Forensic Duplicator
Write-Blocker Kit
Incidence Response Tool for live acquisition of RAM, system
Profile, etc. on site.
Digital Evidence Analysis Tool.
Mobile Phone Forensic analysis tool including SIM Cards
Digital HD Video Camera etc.,
4
Portable
Imaging Device
for Incident
Response
(Encase
Portable
M/S
Guidance
Software)
1. Should have the inbuild blocker device for the source drive.
2. Should support the disks of various interfaces like IDE, SATA,
ESATA, FIREWIRE and USB devices.
3. Should have the facility to Disk to Disk duplication with a facility of
compression.
4. Should have the facility to Disk to File duplication with e01 and
.ex01.‘
5. Should compute and verify the hash values with at least two hashing
algorithms.
6. Should have the facility to format./wipe/black check the destination
drive.
7. Should have the facility to display, print, and save logs.
8. Should have a user-friendly GUI screen.
9. Should have the facility to use the device as a remote write blocker.
5
Forensic Triage
Tool Kit
1. Preview and acquire full disk, targeted data, support Image formats of
AD1, E01, RAW, or SMART.
2. Acquire data from a live system with an active USB port.
3. Manual mode to search the file system/data prior to collection.
4. Pre-configured options for reporting on collected data.
Support to collect data from Chrome Browser History, Default
Browsers, Firefox Browser History, Internet Explorer History,
Internet Explorer Registry Keys, Typed URLs, Recently Accessed
media Player Files, Local Network Connections, Remort Share, IP
Addresses, Application Usage History, Installed Software, Program
Files Software, Start-up Programs, Clipboard Data, Device Drives,
Memory Dump, Processes, Scheduled Tasks, Screenshot, Services,
User Accounts, User Groups, Acquire Registry, System Information,
Typed Paths, USB Devices, Users, Owner Information, SAM User
etc.
Page 64 of 95
6.
Digital Forensic
Tool Kit for the
Analysis of the
Various Internet
Artifacts,
Browsers
(Magnet(IEF)
Internet
Evidence Finder
Magnet Forensic
Inc, Canada)
1. Run directly from the USB thumb drive without having to install the
software on the target computer.
2. Mount & Serch volume shadow copies on a live system, at the logical
or physical level.
3. Save all results directly to the thumb drive.
4. Automatically check for disk encryption including Truecrypt, PGP,
Bitlocker and SafeBoot.
5. Built-in live RAM capture.
6. Should be able to collect/ parse/ carve/ recover artefacts left behind
live/ offline system when using.
Cloud Artefacts line
Dropbox, Carbonite, SkyDrive, Google Docs, Google Drive
Flickr
Social Networking Pages like
Facebook, Twitter, Bebo Chat, Myspace Chat, Google+, Linked
In
Webmail Applications like
Gmail Email, Yahoo Webmail Email, Hotmail Webmail Email
Instant Messenger Chats like
Google Talk Chat Messages, Yahoo Chat Messages, MSN/
Windows Live Messenger Chat Messages, messenger Plus Chat
Logs, AOL Instant Messenger (AIM) chat logs, mIRC Chat logs,
Skype, ICQ, World of Warcraft, Second Life, Trillian
P2P File sharing Application line
Linewire, Frostwire. Props Files, Gigtribe Chat Messages, Ares
P2P Search Keywords, Shareaza Search Keywords, eMules.
Torrent File Artifacts
Web Browser Activity from
Internet Explorer History/Recovery URLs, Firefox History,
Google Chrome History, Apple Safari, Opera, Google Maps,
Browser Activity-Chrome Incognito/Firefox Private Browsing.
7.
Forensic
Software for
the Analysis
of Mac
System (Mac
Forensics Lab
M/S. Mac
Forensics Lab
Inc, USA.)
Analyze Mac OS and dual-boot disk and partition images in multiple
formats
Analyze configuration and log files from common OS applications,
such as Mail, Safari, iChat and Address Book.
Perform rapid searches using metadata.
Gathers comprehensive machine usage information.
List detailed information about every iPod and iPhone that has been
connected to the machine.
Detect VMWare, Virtual Box & Parallels Virtual Machines.
Detect and analyze File Vault-encrypted user directories.
Support dd, EnCase, FTK, AFF, and Apple disk images.
Should run on Microsoft Windows OS as well as Mac OS.
Full support for Mac OS both as an analysis machine and an
investigative target.
Analysis of iCloud configuration Data
Page 65 of 95
Annexure-23
Software/Hardware for analysis of various types of mobile phones,
SIM cards, Memory cards
1
Software Based
Tool Kit for the
Logical Level
Analysis Of
GSM/ CDMA
Mobile Phones
(Universal
Forensic
Extraction
Device (UFED)
Link Analysis)
Should extract phone basic information and SIM Card data.
Should extract phone book and contact list, Call Logs, caller
group information, organizer, notes, SMS messages (live &
deleted). Web-browser artifacts, multi-media messages with
attachments, e-mail messages with attachments, multi-media
files (images, audio & video files), details of installed
applications and their data, traffic and sessions log.
Should support timeline analysis, GEO event positioning
analysis, skype parsing, SQLite database parsing, list parsing.
Should support iPhone backup analysis, blackberry IPD
backup analysis, DMG parsing.
Should Support for parsing the password-protected iPhone
backups.
Should support link analysis.
Should support Social Networking data analysis.
OEM Connectors cables.
2.
Hardware Tool Kit
For The Physical/
Logical Level
Analysis Of
GSM/CDMA
Mobile Phones &
Sim Cards (UFED
Touch 2 M/S.
Cellebrite)
Should Support physical and logical analysis of various mobile
phones working on the various OS platform.
Should Support physical analysis of password-protected,
jailbroken, non-jailbroken, encrypted and non-encrypted iOS
devices.
Able to bypass PIN, passcode lock, pattern lock from the
mobile equipment.
Support to GPS devices.
Able to extract all data (call logs, phone book, SMS messages,
e-mail messages along with attachments, MMS messages along
with attachments, calendar, multi-media files, passwords,
location information, audio and video files etc.)
Able to perform timeline analysis, image carving, conversation
view, SQLite database parsing.
Should support for physical memory dump and file system
dump.
Should support for SIM Card reading, extract SIM related info
along with user information line phone book, call register, text
messages (Active and Deleted ) and SIM Card cloning to
bypass PIN.
Should support for data authentication by hashing
algorithms.
3.
Mobile Phone
Forensic Analysis
Tool Kit For the
Physical Logical
Analysis of Chinese
Mobile Phones
(UFED Chinex
M/S. Cellebrite)
Should support mobile phones working on Chinese chipsets
like MTK, INFINEON, SPREADTRUM, etc.
Should support low-level data extraction from.bin files.
Should support the interpretation of low-level data into a
readable format.
Should recover deleted phone information.
Should recover/ bypass passwords.
Should extract internet artifacts and GPRS activity.
Page 66 of 95
4.
Digital Forensic
Tool Kit for the
Physical Imaging
and Analysis of
SIM Card Memory
(Paraben SIM Card
Seizure Paraben
Corporation, USA)
Read-Only Access to system and user data held on the SIM
card.
Able to handle PIN and PUK.
Present data in an easily readable format.
Authentication by MD5/ SHA hash.
Support for multi-language text messages.
Recover deleted SMS messages.
Page 67 of 95
Annexure-24
Password Cracking Tool for Decrypting the Password Protected
Files
Password Cracking
Tool for Decrypting
the Passwords
Protected Files
(Passware Kit
Forensic)
Recover passwords for various file formats and decrypt encrypted hard
disks.
Should support batch file processing.
Able to scan computer and network for password-protected files and
decrypt the same.
Acquire memory images of the seized computers.
Able to recover User Login passwords.
Able to run from a USB thumb drive and recover password without
installation on a target PC.
Able to reset the password for Local and Domain Windows Administrators
instantly.
Able to recover encryption key for hard disks protected with BitLocker,
including BitLocker ToGo, TrueCrypt, FileVault2 and PGP volumes.
Able to recover password for Windows user from a memory image or a
standalone SAM file, iTunes backups: both iPad and iPhone.
Able to recover passwords from email, websites and network connections
from standalone registry files.
Able to extract passwords from encrypted Mac keychain files.
Page 68 of 95
Annexure-25
High-end configuration Server and Workstations for backup and
analysis
General Specifications:-
Technical
Specifications
for Localized
Server
4U Rack Mounter Server with the following Components:
FILE SERVER:
Multi-core Neon Processor with 8 cores per processor or better
Total RAM Memory 256GB with the provision of scalability to 512GB
Total File Storage Memory of 200TB by RAID6 configuration in SATA
based Secondary Storage Media with provision for future storage expansion.
Support remote acquisition.
Backup support into latest LTO Tapes with backup / archival software.
Cabling with Data Transfer Capability of 10Gbps among file Server and
Processing Server and Work Stations.
Latest Microsoft Windows Server Operating System or suitable OS for the
management of the File Server.
Sufficient Modular Power Supply
Integrated Dual Modular Battery Backup for uninterrupted Power Supply.
Integrated Modular Cooling System for better heat dispensing.
PROCESSING SERVER:
Multi-core Xeon Processor with 10 cores per processor or better
Total RAM Memory 512GB with Provision of scalability.
SSD based Processing Storage of 512 GB with the provision of scalability.
Facility to create custom Virtual Desktops with varying configuration by
visualization
Support to minim 25 clients with scalability.
Support Sharing of Digital Forensic Tools over LAN
Cabling with Data Transfer Capability of 10 GBPS among File Server and
Processing Server and Work Stations.
Latest Microsoft Windows Server Operating System or suitable OS for the
management of the Processing Server.
Sufficient Modular Power Supply
Integrated Dual Modular Battery Backup for uninterrupted Power Supply.
Integrated Modular Cooling System for better heat dispensing.
Integrated GPU supporting the processing of data by Digital Forensic Tools
and decryption related activities.
Integrated Modular Monitor with Keyboard and mouse provision for the
management of the Server via user console management.
The server should have sufficient no. of USB 2/3 ports along with USB C-
port.
Fibre-based LAN to support up to 10 Gbps data speed between server
locations to workstations.
Minimum of 3000 VA rack-mounted integrated dual power supply via UPS.
Warranty for 03 years with lifetime free servicing.
Page 69 of 95
Generalized specifications for workstation:
1. CPU
Intel ® Xeon ® Processor E-5-2650 v2/v3, 2.6 GHz 20 MB L3
cache or higher
2. Chipset and Mother Board
Intel C 602 or better
3. Memory
08 GB DDR3 1600MHz expandable to 256GB or better
4. Hard Disc Drive
1x1000 GB SATA HDD at 7200 rpm upgradable to 2 HDD or
better
5. Storage Controller
SATA Controller RAID Support 0 & 1
6. Key Board
Standard Keyboard
7. Mouse
Optical Scroll Mouse
8. PCI Slot
5 PCI/PCI Express including 2 PCI Express x 16 for Dual
Graphics and TESLA Card support
9. Bays
Total bays (2 internal and 2 external)
10. Ports
5 USB 2.0, 2 USB 3.0, rJ-45, audio in, audio out, mic in
11. Cabinet
Mini tower
12. Optical Drive
8 x DVD writer
13. Networking Features
Integrated 10 / 100/ 1000
14. Operating System
Microsoft windows i.1 Professional 64 Bit or higher
15. OS Certifications
Windows and Red Hat or SUSE Linux or Ubuntu Linux
16. Applications
As per list enclosed at Annexure-A OEM Certification for (a)
Digital Content (b) Electronic Design automation (EDA_ and
(c), Mechanical Computer-aided Design (MCAD) from OEM
Manufacturing workstation shall be acceptable.
17. Safety Certification
As per IEC: 60950-1 / IS:13252 with all amendments up to date
18. Power Supply
230V + / - 10% single phase, 50 Hz AC
19. Power Management
ACPI (Advanced Configuration and Power Management
Interface)
20. Bundled Software
System Health monitoring tool available with H/w box
21. Security
Integrated panel lock or padlock
Page 70 of 95
Annexure- 26
Hardware/Software for Audio Analysis
General Specifications:-
1.
Forensic Audio Examination Hardware and Software System
Capture of Audio
Facility for minimum 4 Channels input or above to collect audio signals simultaneously from
various devices.
Facility for minimum 4 channels output or above, line and speaker, headphone output.
Frequency response for both AC and DC mode 20 Hz 50,000 Hz.
Dynamic range should be greater than 90 dB.
Support the sampling rates 8,000 50,000 Hz or better for analysis of higher frequency signals for
research and case analysis purpose.
Inbuilt features for accurate voice signal capture such as the use of low-latency ASIO drivers, high
gain pre-amplification and anti-alias filtering.
The facility of Software Interface card with Audio Stream Input and Output (ASIO} and Multi-
Media Extensions (MME}.
Facility for segregation of Audio from different formats
Noise reduction of Audio
The facility of Digital Interface card with AES/EBU or S/PDIF connectors and support
various audio formats
Ability to offers signal-to-noise performance typically superior to generic sound cards.
Facility for Signal Normalization
Facility for AC coupling to removes low-frequency components.
Audio features extraction, analysis and display
Facility for waveform, FFT, LPC Spectrograms/spectrums & power spectrum
Facility for Pitch, formant extraction and Cepstrum analysis
Facility for energy and power spectrum
Support the Windows XP or better operating system.
Statistical analysis and display of features
2.
Audio enhancement and authentication Hardware/ software
Capture Uncompressed/compressed audio data from a variety of sources.
Capability to support major audio formats
Audio stabilization.
Facility for pre processing and enhancement of Audio signal using different filters such as
adopting filtering, filter averaging, filter compression and equalization etc.,
Facility to determine the authenticity of the audio
Page 71 of 95
Annexure -27
Hardware/Software for Video Analysis
General Specifications:-
Forensic Image& Video Enhancement and Analysis System
A computerized system with preloaded software.
Capture Uncompressed/compressed audio-data from a variety of video sources
including CCTV.
The system should have editing features such as crop, flip, rotate, re-size,
deinterlace
The system should have the capability to extract and compare the frames.
Pre-processing of the video signal and image clarification.
Video de-multiplexing and stabilization.
Process Digital (DVR) or Analog Video Evidence.
DVR security video-decode digital video (DVR) files from proprietary security
systems into the uncompressed video.
Enhance dark video and poor quality video security and surveillance video.
De-interlace field recorded forensic video to avoid blurry stills printed from video
frame averaging.
Facility for enhancing image/video by Laplacian filtering, Gaussian, Bilateral,
Wiener, Deblocking filters or other suitable filters for forensic applications.
Support for enhancement/Noise suppression (Signal SNR up to 4dB).
Variable slow motion speed adjustment to compensate for time-lapsed video.
The system should be compatible with digital and analogue video, still image and
should be scalable with different modules.
The system should have the facility to analyze with photogrammetric tools and
measurement tools.
Suitable hardware to meet the above requirement.
Page 72 of 95
Annexure-28
Facial recognition Software
General Specifications:-
Facial Capturing from different media & pre-processing.
Creation of photo like composites
Generation of unique alphanumeric code for every composite.
Three-tone hair colour, side to side hair flip.
Data Generation of Facial markings: scars, moles, piercing, tattoos.
Facility for improvement of age progression.
Detachable hats and headwear facility.
Ability to export composite as JPEG file and in other image formats
Side by sideshow I comparison capability.
Improved zooming and positioning tools.
Runs on any standard desk or laptop computer.
Analyze the static Images
Full or partial face identification.
Capable to detect and register the numerous characteristic of each face.
Capable to store, retrieve and compare the facial features from the database and display
matching score.
Page 73 of 95
Annexure-29
Image Analyser Software
General Specifications:-
Should support Windows 10.
Should support JPEG, BMP, TIFF, PNG, RAW image file formats.
The software should have the capability of extracting EXIF metadata of the image.
Capability to of creating camera fingerprint by identifying pixel non-uniformity (PNU) of the
images from a set of test shots which were caused by the fixed pattern noise(FPN) and the
photo-response non-uniformity noise (PRNU) of the sensor.
Variable crop size, pixels and location for PNU identification.
Comparison with a digital image in question and auto-generation of the comparison result.
Page 74 of 95
Annexure -30
Multispectral Imaging system (VSC)
General Specifications:-
Comprising
High-Resolution CCD/ digital FireWire
Color Camera of 1.3 megapixel and above
Camera Spectral Response range 74 approx.. 350 1100 nm
Hyper Spectrometer
Optics
30 flat screen monitor
Magnification up to x170
Field of view 74 approx. 200 x 150 nm
Illumination
Long Wave incident UV at 365 nm
Medium Wave incident UV at 312 / 313 nm
Short Wave incident UV at 254 nm
Long Wave transmitted UV at 365nm
Visible & IR Incident High-Intensity Light
Visible / Infrared Transmitted Flood Light
IR illumination for Anti-Stokes features detection
IR Luminescence
Software (SW)
ICAO data reader
Hidden Image Decoder
Histogram Stretching
Case Work
Side By Side Comparison Of Live & Stored Image
Color Measurement to measure color Coordinates
Comparison of spectrum for whole UV, visible & NIR region
Split Screen for simultaneous ink analysis on two separate documents
Calibration procedure including measuring statistics, image analysis, Processing and
reporting P.C operating system Windows
X-Y translator Stage, Polari safe viewer for UV
Image
Comparison &
Transformation
Superimposition of Live and stored image with the adjustable mix, 0 to 100%
Variable Speed Image storable between a live and a stored image
Left to right image reversal (Horizontal image reversal)
Top to bottom image reversal (Vertical image reversal)
Grey level reversal (Positive to negative)
Image rotation through any angle
Contract / Stretch
Image processing, Image quality enhancement filters, Multi-layer imaging,
IPI decoder, OVI viewer, MRZ decoder
Image
Measurements
Distance between two points
The area within a user-defined box
The radius of a circular feature from three peripheral points
The angle between two lines
Area of user-defined shape
Area of selected features i.e. An alphanumeric character or logo
Data Base
Passport database with a free update for the next 5 years
Bank Note database with a free update for the next 5 years
Computer
Hardware
Compatible Computer hardware for the high-resolution spectral comparator of the
latest version is to be quoted.
Page 75 of 95
Annexure -31
High-resolution Spectral Comparator (HRSC)
General Specifications for High-Resolution-Hyper Spectral Comparator
System
1. High-resolution Hyperspectral comparator System with High-Resolution CCD/digital FireWire
Color Camera with suitable resolution in Megapixels to have a high-quality image to avoid blurry
and pixelated image, Camera Spectral Response range 75 approx.. 400 or less to 1000 rim or
more.
2. The System must have UN illumination facility for examination of UV activated features,
erasures, substrate, fluorescent and phosphorescent inks with Long Wave incident UV (365nm or
more), Long Wave transmitted UV (365nm) light source.
3. The System must have facilities for examination and discrimination of inks, the examination of
erasures, obliterations, additions, deletions, security threads, micro printing, overt and covert
(Active and Passive features) security features in the Passports, Bank Notes, Travel documents;
the system must have Visible & IR incident High-Intensity Light, Visible/Infrared Oblique/side
Light, Visible/Infrared transmitted light, Spotlight source.
4. The System must have visible range light source as well as the provision of transmitted and
oblique light source for examination of watermarks and other surface features.
5. The System must have Visible/Infrared Co-axial light provision for the examination of retro-
reflective security printing features.
6. The System must have facilities for Hyper Spectral Imaging for precise Ink
Examination/discrimination and Image integration.
7. The System must have the facility for examination/viewing and recording of
OVD/DOVD/Holograms.
8. The System must comprise arrays of visible range filters, low and high bandpass filters suitable
for examination of forensic documents.
9. The System must have suitable wide-field IR illumination assembly for
viewing/examination/detection of all Anti-Stokes features.
10. The System must have polarized light for verifying birefringence security features.
11. The System must have ICAO data reader and other ICAO compliant features, Hidden Image
Decoder, Histogram Stretching.
12. The System must have Case Work Management provision.
13. The System must have the facility of Image Comparisons such as side by side comparison of Live
& Stored Image, Superimposition of live & Stored Image, image transformation and Image
annotation provision.
14. The System must have Colour Measurement to measure colour Coordinates.
15. The System must have the facility for Comparison of Spectrums for whole UV, Visible & NIR
region.
16. The System must have the facility of spectrometry for measurement of absorption, reflectance,
transmission and fluorescence spectra.
17. The system must provide Calibration procedure/function.
18. The System must have the provision of full range examination of the document i.e. under various
light sources and suitable magnification with an enhanced field of view along with image storage
and retrieval facility.
19. The System must have UV polarized light viewer for special UV features with UV safety locks
provision.
20. The System must have Variable speed image storable between a live and stored image, Left to
right image Reversal (Horizontal mirror reversal), Top to Bottom image reversal ( Vertical mirror
reversal), Grey level reversal (Positive to negative), Image rotation through any angle,
Contract/Stretch.
21. The System must have the facility of Image processing, Image quality enhancement filters, Multi-
Layer imaging.
22. The System must have IPI/ICI Decoder, OVI Viewer, Latent Image visualizer, Barcode reader
(1D/2D), OCR, Smart card reader, and other formats of Barcodes, MRZ decoder along with
suitable filter light source and software.
Page 76 of 95
23. The System must have the facility of Image measurement facility (software) for measuring the
distance between two points, Area within a user-defined box, Radius of a circular feature from
three peripheral points, Angle between two lines, Area of user-defined shape, Area of selected
features i.e. an alphanumeric character or logo.
24. The System must have Passport, Bank Note and other security documents Database with the
assurance of regular update.
25. The system must have Software to control and run various functions and applications of the
Spectral Comparator during the course of examination as well as handling of the instrument
along with the facility for updating in future.
26. The System must have suitable & efficient hardware and software components to cater & address
the aforesaid forensic needs and mentioned parameters of analysis as well as must-have working
platform/area, sufficient field of view, optical and digital zoom to enable effective & convenient
examination of large-sized documents consisting of extensive data to be examined.
27. The System must have a compatible Computer System with latest windows version and with
latest color printing facility and a minimum 30 flat widescreen monitor or more.
28. Consumable and Spare Parts/Accessories for five years use of equipment as per the Standard
Practice of manufacturer/firm must be provided with the system.
29. Warranty: Three (03) Years.
30. AMC: Supplier shall undertake to enter into AMC of the equipment on payment basis after the
warranty period.
31. The facility of service centre should be available in India.
32. Operational training of seven working days should be provided by the manufacturer for two
scientists at consignee place. Two days of additional training for two scientists in case of
upgradation of Hardware/Software, if any.
Page 77 of 95
Annexure -32
High-End Computer with Server
General Specifications:-
Technical
Specifications for
Localized Server
4U Rack Mounter Server with the following Components:
File Server:
Multi-core Neon Processor with 8 cores per processor or better
Total RAM Memory 256GB with the provision of scalability to 512GB
Total File Storage Memory of 200TB by RAID6 configuration in
SATA based Secondary Storage Media with provision for future storage
expansion.
Support remote acquisition.
Backup support into latest LTO Tapes with backup / archival software.
Cabling with Data Transfer Capability of 10Gbps among file Server
and Processing Server and Work Stations.
Latest Microsoft Windows Server Operating System or suitable OS for
the management of the File Server.
Sufficient Modular Power Supply
Integrated Dual Modular Battery Backup for uninterrupted Power
Supply.
Integrated Modular Cooling System for better heat dispensing.
Processing Server:
Multi-core Xeon Processor with 10 cores per processor or better
Total RAM Memory 512GB with Provision of scalability.
SSD based Processing Storage of 512 GB with the provision of
scalability.
Facility to create custom Virtual Desktops with varying configuration by
visualization
Support to minim 25 clients with scalability.
Support Sharing of Digital Forensic Tools over LAN
Cabling with Data Transfer Capability of 10 GBPS among File Server
and Processing Server and Work Stations.
Latest Microsoft Windows Server Operating System or suitable OS for
the management of the Processing Server.
Sufficient Modular Power Supply
Integrated Dual Modular Battery Backup for uninterrupted Power
Supply.
Integrated Modular Cooling System for better heat dispensing.
Integrated GPU supporting the processing of data by Digital Forensic
Tools and decryption related activities.
Integrated Modular Monitor with Keyboard and mouse provision for
the management of the Server via user console management.
The server should have sufficient no. of USB 2/3 ports along with USB
C-port.
Fibre-based LAN to support up to 10 Gbps data speed between server
locations to workstations.
Minimum of 3000 VA rack-mounted integrated dual power supply via
UPS.
Warranty for 03 years with lifetime free servicing.
Page 78 of 95
Generalized specifications for workstation:
1. CPU
Intel ® Xeon ® Processor E-5-2650 v2/v3, 2.6 GHz 20 MB
L3 cache or higher
2. Chipset and Mother Board
Intel C 602 or better
3. Memory
08 GB DDR3 1600MHz expandable to 256GB or better
4. Hard Disc Drive
1x1000 GB SATA HDD at 7200 rpm upgradable to 2 HDD
or better
5. Storage Controller
SATA Controller RAID Support 0 & 1
6. Key Board
Standard Keyboard
7. Mouse
Optical Scroll Mouse
8. PCI Slot
5 PCI/PCI Express including 2 PCI Express x 16 for Dual
Graphics and TESLA Card support
9. Bays
Total bays (2 internal and 2 external)
10. Ports
5 USB 2.0, 2 USB 3.0, rJ-45, audio in, audio out, mic in
11. Cabinet
Mini tower
12. Optical Drive
8 x DVD writer
13. Networking Features
Integrated 10 / 100/ 1000
14. Operating System
Microsoft windows i.1 Professional 64 Bit or higher
15. OS Certifications
Windows and Red Hat or SUSE Linux or Ubuntu Linux
16. Applications
As per list enclosed at Annexure-A OEM Certification for (a)
Digital Content (b) Electronic Design automation (EDA_ and
(c), Mechanical Computer-aided Design (MCAD) from OEM
Manufacturing workstation shall be acceptable.
17. Safety Certification
As per IEC: 60950-1 / IS:13252 with all amendments up to
date
18. Power Supply
230V + / - 10% single phase, 50 Hz AC
19. Power Management
ACPI (Advanced Configuration and Power Management
Interface)
20. Bundled Software
System Health monitoring tool available with H/w box
21. Security
Integrated panel lock or padlock
Page 79 of 95
Annexure -33
Computerized Polygraph System (CPS)
General Specifications:-
A. Computerized Polygraph system should be based on verified, standard
data/chart scoring software.
1.
The software should have the following applications.
a. System should have Minimum ten(10)channels
b. Automatic & Manual scoring options
c. Event markers and integrated question timing
d. Customizable question templates
e. Licensed & dispute-free polygraph software should be provided
f. Free software upgrades during the warranted period
2.
Physiological Responses Monitoring should have minimum below channels:
a. Two Pneumography (one thoracic & one abdominal)assembly
b. One Galvanic Skin Response (Electro Dermal Response)sensor
c. One standard cardio arm cuff & finger cuff with pump bulb Gauge
d. One Plethysmograph
e. ActivitySensorsforseat,armandfoottomonitorandtorecordphysicalcounter
f. Fingerprint Scanner through the USB port.
g. Digital recording through Webcam
3.
Data Acquisition System interface
4.
Portable Subject Chair
5.
Warrantee of Computerized Polygraph System should be of minimum two(02) years from the
date of installation at the place delivered laboratory.
6.
Instruction Manual of Computerized Polygraph(Hard and soft copy)
B. Computerized Polygraph System should have the following hardware:
a. Latest/compatible Computer configuration is available at the time of delivery.
b. The desktopsystemshouldhaveapre-loadedgenuineAntivirusandWindowsten,PDF writer.
c. Minimum 17-inchMonitor
d. Inkjet Color Printer
Note: The above specification system is available on the GeM portal.
Page 80 of 95
Annexure -34
Layered Voice Analyser
General Specifications:-
Layered Voice Analyzer system should have the following specifications:
Hardware:
Laptop based system with the following specification:
a. Intel i5 processor or higher
b. Windows 10 or higher operation system
c. 1TB Hard Disc or more
d.4GB RAM or more
e. 14‖ inch or more LED display with a carry case for laptop
Headphone with noise-cancelling facility
Microphone with noise cancelling facility
External audio speaker and audio recorder
USB phone connector with an inbuilt sound card
Laser Jet colour printer
Software:
The software should have the following specification:
The software should be based on the psychological pattern of detection on voice samples
Real-time analysis and real-time feedback of the voice input
Option to analyse the pre-recorded voice data
Ability to identify and measure emotional activities and deception from the voice data
The software should have forensic case creation and reporting facility
Licensed and software for dispute free usage of the instrument for an unlimited period
Two-year onsite warranty
Note: Specifications proposed above may vary as modern technology is very dynamic in nature.
Page 81 of 95
Page 82 of 95
GENERAL LABORATORY CONSUMABLES
AND INFRASTRUCTURE REQUIREMENTS
Page 83 of 95
Page 84 of 95
Appendix A
General Laboratory Accessories
Refrigerator (explosion-proof)
Oven
Fume hood (for handling toxic and flammable chemicals)
Biohazard hood (for handling bio hazardous materials)
Hot plates or heating mantles
Ultrasonic bath
Mortars and pestles
Spot test plates
Test tube stands
Test tubes and corks (or screw-cap test tubes)
Supports (rectangular base with clamps, and ring support with clamps)
Tubing (regular and thick-walled for vacuum)
Magnetic stirrer and stirring bars
Timer/clock
Silicone grease
Bunsen burner (with suitable gas supplycan be portable butane/LPG)
pH indicator paper, 0-14
Swab boxes
Paper towels
Sealing Wax
Glassware
Beakers, assorted from 5 to 1000 ml
Erlenmeyer flasks, assorted from 10 to 2000 ml
Test tubes, length 130 mm, diameter 25 mm
Desiccators with the disc, vacuum, safety, with screw thread connection stopcock in the lid,
top ID 250 mm
Desiccant
Glass tubing
Glass rods
Dispensing flasks fitted with eye-droppers
Glass stoppers, assorted
Reagent flasks
Pasteur pipettes
Teats for Pasteur pipettes
Reagent labels
Volumetric measurements
Graduated cylinders, assorted from 5 to 1000 ml
Graduated pipettes, assorted from 1 to 10ml
Volumetric pipettes, assorted from 1 to 25 ml
Volumetric flasks, assorted from 5 to 1000 ml
Page 85 of 95
Weight measurements
Analytical balance, 0.1 mg resolution with internal calibration
Top-pan balance, 0.01g resolution
Calibration weights
Weighing dishes (e.g. disposable plastic boats)
Solution preparation
Beakers, assorted from 10 to 1000 ml
Conical flasks, assorted from 50 to 2000 ml
Graduated cylinders, assorted from 5 to 1000 ml
Electric hot plate/magnetic stirrer unit
Magnetic stirrer bars (assorted)
Extraction and separation
Mechanical shaker
Centrifuge
Rotary beater
Soxhlet apparatus
Thimbles
Separator funnels 50, 100 and 200 ml
Funnels for gravity filtration, diameter 40-60 mm
Buchner funnels for vacuum filtration, diameter 65-85 mm
Suction rubber adaptor set
Centrifuge tubes
Filter paper, assorted from 7 to 12.5 cm
Large sheet of filter paper and Bench rolls
Distillation and evaporation
Rotary evaporator
Water distillation apparatus for the distillation of tap water including a storage container
Diagonal condenser including a receiving and an evaporating/rotating flask (vol. 1L)
Round bottom flasks, assorted from 50 to 1000 ml
Supports for round bottom flask
Laboratory jacks
Water bath
Clamps
Vortex
Vacuum pump
Recirculation chiller
Disposable face masks in addition to lab coats
Bio-waste containers
Sharps disposal containers
Appropriate buffers for extraction technique
Cleaning and decontamination reagents such as appropriate preparations of
bleach and alcohol
Page 86 of 95
For Thin Layer Chromatography
Glass or aluminium silica gel coated plates (20x20, 20x10, 20x5 cm), with and
without a fluorescent indicator.
Developing tanks
TLC spray box (a cardboard box can be used)
Atomizers with rubber bellows for spray reagents
Air blower (to dry plates after sample application or after development)
The ultraviolet lamp (254 and 366 nm)
Desiccators for TLC plate storage
Solvents
Spray reagents (colour reagents/iodine vapours)
Glass micro capillary tubes, 2 μl
Filter paper (for development chamber saturation)
Multipurpose spotting guide, 20x20 cm
Reference standards (known compounds)
TLC plate heater
If no commercial silica gel plates are available
Spreading table with levelling device
TLC plate coater/Applicator
Drying and/or storage rack for 20x20, 20x10 and 20x5 cm TLC plates (temperature
range 40-250°C)
Glass plates 20x20, 20x10, 20x5 cm (86 approx.. 4mm thickness)
Silica gel with and without fluorescent indicator
Page 87 of 95
Appendix B
For Equipment
Special gases facilities (gas cylinders with regulators, valves, molecular sieves tubes)
Data processing facilities
Crimp sealer
Decapper
Stainless steel tweezers
Wrenches
Vial racks
Glass vials
High purity (5.0) gases (H2 , N2, Helium, aitr)
Derivatization reagents (e.g., MSTFA, BSTFA, TMCS)optional
Hot plate/heat block or water bath to warm-up vials.
Capillary column cutter
Reference/standard substances and test mixtures
De-ionized water
Specific reagents for buffers and mobile phases
Syringes, 1-3 ml (for sample filtration)
Syringes filters with the adequate membranes (e.g. 30mm diameter, membrane
o PTFE, porosity 0.45 μm.
KBr die (e.g., 13 mm diameter)for use in halide disc method
Micro KBr die (e.g., 1.5-3 mm diameter)for use in halide disc method
Sample holder
Demountable cell mounts
Agate mortar (e.g., 35 mm diameter) and pestle
Teflon spacers
Stoppers
Paper rings (e.g., for 13 mm die)for use in halide disc method
Syringe, 2 mlfor use in Nujol mull method
Nujolfor use in Nujol mull method
KBr powderfor use in halide disc method
Liquid nitrogen
For general physical measurements (including photography)
Computer, including CD burner and office application software
Scanner
Printer
LAN with Internet access
High-End SLR camera.
Photographic software e.g. Adobe Photoshop®
Rulers (1 metre)
Tape measure with larger-scale increments for higher visibility in photographs e.g.
2-metre x 25mm mini measuring tape Pocket Style
Assorted size reference scales (both solid plastic and adhesive disposable)
Elasticised ―hi-visibility‖ builders string or similar
Strapping tape or other re-enforced adhesive tape
Bulletproof equipment such as vests, Helmets, goggles etc.
Scientific calculator
Zero base-line protractor
Page 88 of 95
Spare camera batteries
Range of lenses (normal, wide-angle and macro)
Lens brush and lens tissue
Image storage cards (e.g. compact flashcards)
Photographic tripod stand
Light meter
Flashlight (and batteries)
Copy stand, with supports for evidential material
Filters
Spare camera batteries and/or charger
Rechargeable hand-held forensic light source with UV and IR capability (with
o associated coloured and polarized photographic filters and goggles)
Software for analysis of bloodstain pattern.
Hand-held metal detector
Dental stone and plaster of Paris for shoe and tyre prints,
Bucket, bowls and/or plastic bags for mixing
Wooden spatula
Wire or wooden splints for support, or casting forms
Silicone casting material for tool marks and indented fingerprints [tube]
For DNA Analysis
Portable autoclave
Dissecting microscope
Surgical Instrument set
Autopsy saw
Pipette sets of different measurements.
Hot plate stirrer
Microwave
Drying cabinet
Page 89 of 95
Appendix-C
For Digital Forensics
Laboratory forensic workstation with the monitor, output devices (CD/DVD) and connections
(i.e., USB, Firewire, SATA) configured with the operating system (i.e., Windows or Linux) and
user-based software applications (i.e., Office, anti-virus); special configuration considerations
include a high-end processor, a significant amount of random access memory(RAM), high-quality
video and audio components, DVD and CD drives, multiple peripheral connection capabilities
for IDE, SCSI, SATA, USB, Firewire, and networks.
Laptop or portable forensic workstation with similar configuration and capabilities to those of
the laboratory forensic workstation
Uninterruptible power supply (UPS) for backup power
Isolation transformer (power/voltage regulation) for PC workstations and audio/video
equipment
Networking equipment
Internet connectivity for updating software and research
General use printer
Digital camera
Cables and connectors
Power surge protectors/extension cords of various lengths
Hard drives
Compact flash cards (various sizes and volumes)
Small computer instrument toolkit
Anti-static wrist band
RF shielding equipment for mobile telephone examinations
High-resolution monitors
Professional quality player of various media formats and DVD recorders.
High-quality video cables appropriate to the equipment in the lab: analogue video cables, digital
video cables
Image printers
Colour calibration equipment for monitors and printers
Page 90 of 95
Appendix D
List Of Minor Equipment/Tools Required For Searching Scene of
Occurrence
i.
For Protection and safety
Disposable latex or plastic examination gloves
Disposable coveralls/Aprons
Hair caps
Shoe covers
List of hazardous chemicals signs
Heavy gloves
Crime scene barrier tape
Disposable face masks (dust and anti-putrefaction masks)
First aid kit
ii.
Detection
Magnifier
Rechargeable flashlight/torch/hand-held light source
Rechargeable hand-held forensic light source (basic)
Goggles
iii.
Collection of Physical Evidence
Clear adhesive tape
Disposable spatulas
Scissors
Scalpels
Disposable plastic tweezers
Scalpel replacement blades
Screw-cap test tubes
Individually wrapped sterile cotton swabs
Disposable plastic pipettes
iv.
For Packaging
Cardboard boxes
Metal unlined cans for collection of arson evidence
Bindle material/weighing paper
Plastic bags
Paper bags (small, medium, large)
Slide boxes (for teeth, projectiles etc.)
Envelopes (various sizes)
Body bags
Containers for sharp objects (e.g. knives, syringes)
Evidence tags/labels
Tamper-proof evidence tape
v.
For Documentation (including photography)
Rulers (30 cm)
Page 91 of 95
Reference scales (L-shaped)
Tape measure
Writing pad
Clipboard
Writing and marking pens, pencils, metal scribe, chalk, marking paint
Paper towel
Stapler and staples
Compact camera
spare batteries, lens brush, lens tissue, image storage cards, filters), memory
card
Compass
Photographic tripod stand with water-level
Camera flash (and batteries)
Magnet
Compass
GPS Locator
Stationery items such as pens, markers, sketch pens, different papers, Chart Papers etc
vi.
Tool kit
Hammers, saw, screwdrivers, wrench, pliers, knife, shovels, sifters, rake, bolt
cutters, power drill, electrical extension cords, wire cutters, hacksaw, socket wrench
set, rope, assorted sized of wood chisels, axe, cotton work gloves etc.
*****
Page 92 of 95